Evaluation of CyberKnife® Fiducial Tracking Limitations to Assist Targeting Accuracy: A Phantom Study with Fiducial Displacement

Abstract: Introduction The underlying assumptions of the CyberKnife® (Accuray, Sunnyvale, CA, US) fiducial tracking system are: i) fiducial positions are accurately detected; ii) inter-fiducial geometry remains consistent (rigid); iii) inter-fiducial geometric array changes are detected and either accommodated with corrections or treatment is interrupted. However: i) soft-tissue targets are deformable & fiducial migration is possible; ii) the accuracy of the tracking system has not previously been examined with fiducial… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
15
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 12 publications
(15 citation statements)
references
References 26 publications
0
15
0
Order By: Relevance
“…During the CyberKnife treatment of prostate cancer, the fiducial tracking system monitors fiducial markers near the treatment site. This system calculates the offset values and corrects it when the marker array differs between live images and digitally reconstructed radiographs (DRR) images in the TLS 14 . For the fiducial tracking system, the offset values (mm) of the three axes (X‐axis; superior [+] to inferior [−], Y‐axis; right [+] to left [−], and Z‐axis; posterior [+] to anterior [−]) can be obtained by using one or two fiducial markers, although the correction for the other axis (pitch, roll, and yaw [degrees]) needs a minimum of three fiducial markers.…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations
“…During the CyberKnife treatment of prostate cancer, the fiducial tracking system monitors fiducial markers near the treatment site. This system calculates the offset values and corrects it when the marker array differs between live images and digitally reconstructed radiographs (DRR) images in the TLS 14 . For the fiducial tracking system, the offset values (mm) of the three axes (X‐axis; superior [+] to inferior [−], Y‐axis; right [+] to left [−], and Z‐axis; posterior [+] to anterior [−]) can be obtained by using one or two fiducial markers, although the correction for the other axis (pitch, roll, and yaw [degrees]) needs a minimum of three fiducial markers.…”
Section: Methodsmentioning
confidence: 99%
“…Among the several radiotherapy modalities, CyberKnife (Accuray, Sunnyvale, CA) can be used to acquire 2D images per 5-150 s by using 2-kV x-ray devices (target locating system [TLS]), resulting in high-precision image-guided radiotherapy. 14 Prostate SBRT using CyberKnife showed fewer grade 2 or worse genitourinary toxicities than other SBRT delivery methods. 15 However, larger intrafractional motion was observed in CyberKnife treatment than in conventional Linac treatment, because the irradiation time for CyberKnife treatment was likely to be long.…”
mentioning
confidence: 88%
See 2 more Smart Citations
“… 3D and 4D imaging (cone beam CT-CBCT).  Newer technologies including imaging and monitoring of electromagnetic transponders (55)(56)(57) or small fiducial markers inserted near the tumor and visible on X-rays or CBCT (58)(59)(60)(61)(62).  Lately surface guidance is used for verification and monitoring of the patient set-up (63,64).…”
Section: Verification Of the Patient Positioning And Target Locationmentioning
confidence: 99%