2024
DOI: 10.1002/eej.23471
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of a vacuum insulated temperature calibration device to calibrate thermistor characteristics for skin cancer diagnosis

Homei Fujita,
Takashi Nonaka,
Yasushi Hosokawa
et al.

Abstract: Recent methods using thermistors as an early detection of skin cancer require the higher resolution of 5 mK. Previously, the authors developed a vacuum insulated temperature device that calibrates the resistance temperature characteristics of thermistors, so that it improves the temperature resolution and accuracy of skin cancer diagnosis. In this study, the operational methods of the calibration system were compared for improving further accuracy using the temperature calibration device and the oil tank calib… Show more