Ep-1440 Dosimetry Comparison of Vmat Dose Delivery Capability of Beam Matched Linear Accelerators in Multi Site Environment

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
7
0

Year Published

2019
2019
2022
2022

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(7 citation statements)
references
References 0 publications
0
7
0
Order By: Relevance
“…In the literature, the quality or accuracy of beam matching has been assessed by the standard deviation in gamma pass rates for a particular treatment, with standard deviations around 1% constituting excellent agreement. Accordingly, this study has shown that the linacs considered are also well matched in parameters (output factors, depth‐dose curves, profiles, energy and dose rate, MLC positional accuracy, MLC speed, jaw positional accuracy, gantry speed etc …”
Section: Discussionmentioning
confidence: 81%
See 1 more Smart Citation
“…In the literature, the quality or accuracy of beam matching has been assessed by the standard deviation in gamma pass rates for a particular treatment, with standard deviations around 1% constituting excellent agreement. Accordingly, this study has shown that the linacs considered are also well matched in parameters (output factors, depth‐dose curves, profiles, energy and dose rate, MLC positional accuracy, MLC speed, jaw positional accuracy, gantry speed etc …”
Section: Discussionmentioning
confidence: 81%
“…demonstrated that a single TPS model could not be supported by their three Elekta linacs which were matched according to the vendor's beam matching criteria, calling for stricter criteria and an additional subset of dosimetric data to be matched to if a single model was to be used. Alternatively, a study by Swamy et al., like Sjöström et al., also considered “fine beam‐matched” Varian linacs and found that from delivery of fifteen volumetric modulated arc therapy (VMAT) plans, the standard deviation in gamma (2%/2 mm) pass rates was 1.00%, demonstrating excellent beam matching in terms of VMAT delivery . To date there have not been any studies published considering beam matching for stereotactic body radiotherapy (SBRT) and the tighter tolerances and objectives it might require, so the feasibility of distributive QA and treatment delivery of SBRT should thus be assessed and verified.…”
Section: Introductionmentioning
confidence: 99%
“…The ability to match linacs at a level sufficient for routine treatments and PSQA has been assessed in several studies 1–8 . Initial investigations focus on the ability to achieve a high level of matching between characteristics of the linac output, such as beam profiles, depth dose curves, and output factors 1,2,4 .…”
Section: Introductionmentioning
confidence: 99%
“…The ability to match linacs at a level sufficient for routine treatments and PSQA has been assessed in several studies. 1 , 2 , 3 , 4 , 5 , 6 , 7 , 8 Initial investigations focus on the ability to achieve a high level of matching between characteristics of the linac output, such as beam profiles, depth dose curves, and output factors. 1 , 2 , 4 Further investigations consider the resultant similarity within composite dose deliveries for volumetric‐modulated arc therapy (VMAT) or intensity‐modulated radiotherapy (IMRT).…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation