Enhanced Ion Mobility Separation and Characterization of Isomeric Phosphatidylcholines Using Absorption Mode Fourier Transform Multiplexing and Ultraviolet Photodissociation Mass Spectrometry

Abstract: The structural diversity of phospholipids plays a critical role in cellular membrane dynamics, energy storage, and cellular signaling. Despite its importance, the extent of this diversity has only recently come into focus, largely owing to advances in separation science and mass spectrometry methodology and instrumentation. Characterization of glycerophospholipid (GP) isomers differing only in their acyl chain configurations and locations of carbon–carbon double bonds (CC) remains challenging due to the need … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
27
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
6

Relationship

3
3

Authors

Journals

citations
Cited by 28 publications
(33 citation statements)
references
References 55 publications
0
27
0
Order By: Relevance
“…FT-IMS acquisitions commonly require ≥5 minutes sweeps to adequately sample CCS distributions. ,,,, Shorter acquisitions down to 1 min were recently shown to be possible by artificially extending the XICs (Figure S1b) using zero-padding and apodization (Figure S1c). Here, the 8–10 min sweeps collected for all proteins listed in Table S1 were truncated to 1 min and 3 min and reprocessed using the aFT method as described previously.…”
Section: Resultsmentioning
confidence: 99%
See 3 more Smart Citations
“…FT-IMS acquisitions commonly require ≥5 minutes sweeps to adequately sample CCS distributions. ,,,, Shorter acquisitions down to 1 min were recently shown to be possible by artificially extending the XICs (Figure S1b) using zero-padding and apodization (Figure S1c). Here, the 8–10 min sweeps collected for all proteins listed in Table S1 were truncated to 1 min and 3 min and reprocessed using the aFT method as described previously.…”
Section: Resultsmentioning
confidence: 99%
“…The PCB-DT with a 10 cm desolvation region and a 10 cm drift region was assembled as previously described. ,, A schematic representation of the instrument coupled to a prototype Thermo Scientific Q Exactive Plus UHMR Orbitrap mass spectrometer (Bremen, Germany) is shown in Figure S1a, and FT-IMS parameters are listed in Table S1. Further details about the modular drift tube can be found in the Supporting Information.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations
“…All data were collected using a modified, research-grade Thermo Scientific Q Exactive HF-X Biopharma mass spectrometer (Bremen, Germany) that was modified by the addition of a 500 Hz, 193 nm coherent ExciStar excimer laser (Santa Cruz, CA) to perform UVPD in the HCD collision cell, as described previously. , For all direction infusion experiments, an nESI NanoFlex source was utilized. The same source was used for interfacing of a nano-LC column eluent into the mass spectrometer.…”
Section: Methodsmentioning
confidence: 99%