Energy and angular dependence of the personal dosemeter in use at ITN-DPRSN

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2011
2011
2021
2021

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 7 publications
(3 citation statements)
references
References 6 publications
0
3
0
Order By: Relevance
“…Dosimeters were read using two Harshaw 6600 readers. The extremity dosimeters were calibrated in terms of H p (0.07) using a N120 X‐ray beam incident on an ISO rod phantom, and the WB dosimeters were calibrated in terms H p (10) and H p (0.07) using a 137 Cs beam incident on a ISO water phantom . The minimum limit of detection is 0.02 mSv in terms of H p (10) and H p (0.07) for WB detectors, and 0.07 mSv in terms of H p (0.07) for the extremity detectors.…”
Section: Methodsmentioning
confidence: 99%
“…Dosimeters were read using two Harshaw 6600 readers. The extremity dosimeters were calibrated in terms of H p (0.07) using a N120 X‐ray beam incident on an ISO rod phantom, and the WB dosimeters were calibrated in terms H p (10) and H p (0.07) using a 137 Cs beam incident on a ISO water phantom . The minimum limit of detection is 0.02 mSv in terms of H p (10) and H p (0.07) for WB detectors, and 0.07 mSv in terms of H p (0.07) for the extremity detectors.…”
Section: Methodsmentioning
confidence: 99%
“…Hp(0,07) is defined as the dose equivalent to soft tissue at a depth 0.07 mm below a specified point on the body. The TLDs were calibrated in terms of Hp(0.07) using a N120 X‐ray beam incident on an ISO rod phantom (17–20) . The limit of detection is 0.07 mSv in terms of Hp(0.07).…”
Section: Methodsmentioning
confidence: 99%
“…Some well-known reference sources used in the calibration procedure are Cs-137, Co-60, and Sr-90/Y-90 (International Commission on Radiation Units and Measurements 1992, Szumska et al 2013, Wrzesień 2020. Moreover, calibrations are also performed with the use of low-energy x-rays (Alves et al 2008, Borges et al 2014, Wrzesień 2019, Kr-85, or using the self-irradiation phenomenon of detectors made of potassium halide (Pashchenko et al 1995).…”
Section: Introductionmentioning
confidence: 99%