Effects of Waveform Model on Sensitivity Values of Transducers Used in Mechanical Dynamic Measurements

Abstract: Dynamic calibration of the pressure transducers and accelerometers are carried out by applying dynamic mechanical inputs to them. Determination of those transducers' sensitivities, defined as the ratio of electrical output of the transducer to the mechanical input, is an important task for calibration laboratories. Data obtained during calibration are processed in order to have the peak values of the input and output signals which are sampled by data acquisition boards. Different approximations are made such a… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
5
0
1

Year Published

2017
2017
2023
2023

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 9 publications
(6 citation statements)
references
References 1 publication
0
5
0
1
Order By: Relevance
“…The dynamic calibration machine is used to create a half-sinus signal with a width of around 5 ms for calibration of dynamic pressure transducers. The measurement range is from 50 MPa to 500 MPa in hydraulic media with targeted relative measurement uncertainty is 1 % [1].…”
Section: Working Principle Of Dynamic Calibration Machinementioning
confidence: 99%
See 2 more Smart Citations
“…The dynamic calibration machine is used to create a half-sinus signal with a width of around 5 ms for calibration of dynamic pressure transducers. The measurement range is from 50 MPa to 500 MPa in hydraulic media with targeted relative measurement uncertainty is 1 % [1].…”
Section: Working Principle Of Dynamic Calibration Machinementioning
confidence: 99%
“…The piston is directly located in line the vertical axis of free dropped mass. (Yasin Durgut et al, 2015). Produced charge values by hitting effect of the mass on to the piston are carried through the green charge cables to channel 1 and channel 2 of the Signal Conditioner/Amplifier Kistler Type 6907B.…”
Section: Performance Testmentioning
confidence: 99%
See 1 more Smart Citation
“…In drop hammer/mass system configuration, test transducer is mounted to close part of a piston and cylinder unit in a closed reservoir or chamber. Piston-cylinder also [38,39] mounted into a reservoir which is filled with fluid too. Fluid is practically incompressible oil like glycerine.…”
Section: Literature Reviewmentioning
confidence: 99%
“…Application of ISO 16063-21 and 22 was introduced by some of the manufacturers [5]. Dynamic calibration of different transducers, such as force and pressure transducers, were investigated * e-mail: eyup.bilgic@tubitak.gov.tr and the results were published, taking care of the applied methods and also of the effects of waveform model [6,7]. Pulse shape effect, uncertainties of the calibrations and also filtering of the noise were studied by various authors [8][9][10].…”
Section: Introductionmentioning
confidence: 99%