2022
DOI: 10.1002/jms.4877
|View full text |Cite
|
Sign up to set email alerts
|

Effects of digital processing on repeatability assessment of a multiple reaction monitoring liquid chromatography–tandem mass spectrometry system by ISO 11843‐7

Abstract: ISO 11843 part 7 (ISO 11843-7) can provide a standard deviation (SD) of area measurements of a target peak through the stochastic behaviors of instrumental noises. The purpose of this study is to demonstrate that ISO 11843-7 can be applied to assess repeatability in an isocratic liquid chromatography-tandem mass spectrometry (LC-MS/MS) system without repetitive measurements. The relative standard deviation (RSD) of the peak area of ergosterol picolinyl ester, which was used as an example, on a multiple reactio… Show more