Effect of Syringe Design on the Accuracy and Precision of Intravitreal Injections of Anti-VEGF Agents

Abstract: Intravitreal injections of anti-VEGF agents using a 1.0 mL tuberculin syringe demonstrate poor accuracy and precision. A lower capacity syringe may improve accuracy, while a low dead space plunger may improve precision.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
13
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 18 publications
(13 citation statements)
references
References 11 publications
0
13
0
Order By: Relevance
“…Still, the chosen method is comparable to the clinical setting; drugs must be withdrawn from the vial in order to be injected into the eye, irrespective of whether the label recommendations or pharmaceutical compounding is utilized. A limitation of the current method itself is that the plunger resistance is altered 38,39 . For the 1.0 mL Injekt-F syringe used in the present study, there were only minimal dose variations.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…Still, the chosen method is comparable to the clinical setting; drugs must be withdrawn from the vial in order to be injected into the eye, irrespective of whether the label recommendations or pharmaceutical compounding is utilized. A limitation of the current method itself is that the plunger resistance is altered 38,39 . For the 1.0 mL Injekt-F syringe used in the present study, there were only minimal dose variations.…”
Section: Discussionmentioning
confidence: 99%
“…For the 1.0 mL Injekt-F syringe used in the present study, there were only minimal dose variations. Yet, previous studies have shown highly variable accuracy and reproducibility when utilizing syringes that are commonly used for intravitreal injections 3942 . Thus, a silicone oil-free syringe with a lower volume specifically manufactured for intravitreal use could further improve the current method.…”
Section: Discussionmentioning
confidence: 99%
“…Moisseiev et al were able to demonstrate that different types of syringes, despite having the same labelling, deliver different volumes and thus deviate significantly from the values that the surgeon would like to administer during the injection. 52 The three syringe models tested in this study injected a significantly higher volume than the intended 50 μL. In the study described, the syringe design with low dead space showed the highest reproducibility of the delivered volume.…”
Section: Options To Reduce Contamination and Side Effectsmentioning
confidence: 75%
“…In a study by Moisseiev et al, 11 100 to 150 μL fluid was pulled into the injectors for IVI, the plunger was pressed up to the 50 μL line, and the fluid was ejected. The mean amount of the ejected fluid was found to be 58.0 ± 5.7 μL (a range of 44.1–70.2 μL) in 1 mL of tuberculin injectors.…”
Section: Discussionmentioning
confidence: 99%
“…1,710 Due to the lack of availability of specific injectors or prefilled ones, 1 cc insulin injectors are generally used in intravitreal injection procedures. 11 The doses of the drugs given in IVI may not be the intended doses. Significant deflections were reported in accuracy, sensitivity, and repeatability of the dose because of the administration of small volumes using relatively large injectors in IVI.…”
Section: Introductionmentioning
confidence: 99%