alfunctioning of a pacemaker because of electromagnetic interference (EMI) was reported after development of the demand pacemaker. 1 Clinically, a variety of medical appliances that produce electromagnetic or radiofrequency waves are now used for diagnostic or therapeutic purposes, and precautions or restrictions on their use have been enforced to protect patients with an implanted pacemaker from possible hazards. [2][3][4][5][6] Computed tomography (CT) is widely used in clinical practice, but a detailed report has not been published of the effects of CT scanning on the operation of pacemakers. Although it is commonly believed that CT scanning does not affect the functioning of pacemakers, we have previously reported a transient malfunction of the pacemaker probably caused by over-sensing. 7 The present study was undertaken to examine the influence of multislice spiral CT on pacemakers in patients and experimental models of the human body.
Methods
ECG Monitoring During Chest CT Scanning in Pacemaker-Implanted PatientsIn 11 patients with implanted demand-type pacemaker, chest CT scanning using a multislice spiral CT system (4-detector row, SOMATOM Volume Zoom, Siemens, Germany) was performed for further evaluation of abnormal shadows that had been observed on chest X-ray.During the CT scanning, the ECGs of the extremity leads were recorded at a paper feed rate of 25 mm/s. All patients gave informed consent.
Measurement of Alternating Electric and Magnetic Fields in the CT Room and on CT Scan LinesThis experiment was conducted to examine whether the CT procedure produces electromagnetic fields in the CT room and induces EMI with pacemakers. Alternations of the electric and magnetic fields on the CT scanning line, and at points 1 m distant from the CT scanning line within the CT room, were measured during CT scanning using an electric-field measuring device (FD-1, Combinova, Sweden) and a magnetic-field measuring device (Model 5080, F.W. Bell, Orlando, FL, USA). The measurements were repeated 3 times, and the maximum values were adopted.
Effects of CT Scanning on Pacemaker Function in Human Body ModelsThe pacemakers, Thera SR8960i and Kappa SR701 (Medtronic, Minneapolis, MN, USA), combined with a 5024M lead (Medtronic) were mounted in Irnich's human body model (Fig 1) 8-10 and subjected to CT scanning. The bench model (Fig 2) was constructed for this study to clarify the conditions induced by CT scanning that might influence pacemaker function.Measuring Systems The Irnich's body model was filled with 0.18 w-% saline, and the electroconductivity of the model was set at a value equivalent to that of the human body. Pick-up electrodes to receive pacing pulses from the pacemaker were attached to the pacemaker lead. Specific electrodes to input the sensing signals (known as cenelic patterns) produced by the pseudo-beat generator were also arranged on the model. The model was connected to the pseudo-beat generator and recorder via a 20-m cable that Circ J 2006; 70: 190 -197 (Received August 24, 2005; revised manuscrip...