2020
DOI: 10.1002/acm2.13106
|View full text |Cite
|
Sign up to set email alerts
|

Effect of an integral quality monitor on 4‐, 6‐, 10‐MV, and 6‐MV flattening filter‐free photon beams

Abstract: PurposeTo investigate the effect of an integral quality monitor (IQM; iRT Systems GmbH, Koblenz, Germany) on 4, 6, 10, and 6‐MV flattening filter‐free (FFF) photon beams.MethodsWe assessed surface dose, PDD20,10, TPR20,10, PDD curves, inline and crossline profiles, transmission factor, and output factor with and without the IQM. PDD, transmission factor, and output factor were measured for square fields of 3, 5, 10, 15, 20, 25, and 30 cm and profiles were performed for square fields of 3, 5, 10, 20, and 30 cm … Show more