Early Implant Failure of a 2-Level M6-Cervical Total Disc Replacement

Abstract: Case: We report a case of implant failure after primary 2-level M6-cervical disc replacement (M6-C; Spinal Kinetics) at C4-5 and C5-6. At revision surgery, a full-thickness tear was identified in the sheath of the C5-6 implant. Histology from periprosthetic tissue demonstrated polyethylene particles without acute inflammation. Tissue cultures were positive for Propionibacterium acnes, but ultrasonicate cultures were negative for both prostheses. An independent laboratory test confirmed mechanical f… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

1
16
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 11 publications
(17 citation statements)
references
References 27 publications
1
16
0
Order By: Relevance
“…However, in several studies in the present review, the early onset of bone resorption led authors to conclude that wear debris was an unlikely primary cause [39,41,44,45,53]. In late onset cases, progressive osteolysis was attributed to mechanical failure of the M6-C in five studies, despite findings consistent with device failure and histopathology confirming the presence of particulate polymeric debris and giant cells [29,30,33,35,36]. This device was noted to include a sheath intended to contain and decrease debris migration, and rupture of the sheath contributed to particle-induced osteolysis [35,36].…”
Section: Wear Debrismentioning
confidence: 68%
See 4 more Smart Citations
“…However, in several studies in the present review, the early onset of bone resorption led authors to conclude that wear debris was an unlikely primary cause [39,41,44,45,53]. In late onset cases, progressive osteolysis was attributed to mechanical failure of the M6-C in five studies, despite findings consistent with device failure and histopathology confirming the presence of particulate polymeric debris and giant cells [29,30,33,35,36]. This device was noted to include a sheath intended to contain and decrease debris migration, and rupture of the sheath contributed to particle-induced osteolysis [35,36].…”
Section: Wear Debrismentioning
confidence: 68%
“…Duration of implantation prior to reoperation ranged from 15-96 months. Radiographic analysis was performed at multiple time points, with progressive osteolysis observed in eight articles [26][27][28][29][30][31][32]. In twelve articles, the implants were removed.…”
Section: Reported Osteolysismentioning
confidence: 99%
See 3 more Smart Citations