“…of triggering by electromagnetic interference in integrated bipolar leads (<1/400 versus 1/80 patient-years). 11,26,32 Once an alert is triggered, the lead integrity alert sounds an audible tone, initiates an immediate remote-monitoring notification in ICDs with wireless telemetry, and reprograms the number of intervals to detect VF to 30/40, which reduces inappropriate shocks caused by transient, fracture-induced oversensing. 11 In a prospective study, the lead integrity alert reduced inappropriate shocks compared with conventional impedance monitoring (lead integrity alert 38% versus control 70%, P<0.001).…”