Abstract: EBT3 film pieces inserted in between the 3 mm sheets. All the measurements were compared with the results from the GammaPlan calculations using both the convolution algorithm and the TMR10 algorithm. Results: The mean, minimum, maximum differences between the ion chamber measured doses from the Tg 21 formalism and the calculated doses from the TMR algorithm are-5.74%,-4.65%,-8.43% respectively. The mean, minimum, maximum differences between the measured doses from the Tg 21 formalism and the calculated doses f… Show more