Abstract:EBT3 film pieces inserted in between the 3 mm sheets. All the measurements were compared with the results from the GammaPlan calculations using both the convolution algorithm and the TMR10 algorithm. Results: The mean, minimum, maximum differences between the ion chamber measured doses from the Tg 21 formalism and the calculated doses from the TMR algorithm are-5.74%,-4.65%,-8.43% respectively. The mean, minimum, maximum differences between the measured doses from the Tg 21 formalism and the calculated doses f… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.