2021
DOI: 10.1177/15330338211011964
|View full text |Cite
|
Sign up to set email alerts
|

Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top

Abstract: Purpose: To evaluate the dosimetric accuracy of the default couch model of the QFix kVueTM Calypso couch top in the treatment planning system. Methods: With the gantry 180°, field size 20 × 20 cm, 6 MV, we measured the depth dose, off-axis dose, and dose plane of different depths in the phantom with the couch rails in and out, respectively. Isocenter doses at different angles were also obtained. The results were compared to the doses calculated using the default couch top model and the real scanned couch top m… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

1
0
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(1 citation statement)
references
References 35 publications
(46 reference statements)
1
0
0
Order By: Relevance
“…Moreover, the prostate VMAT plans were not significantly affected by the introduction of this device. Our results are consistent with other studies [9][10][11].…”
Section: Discussionsupporting
confidence: 94%
“…Moreover, the prostate VMAT plans were not significantly affected by the introduction of this device. Our results are consistent with other studies [9][10][11].…”
Section: Discussionsupporting
confidence: 94%