Dosimetric comparison of AcurosBV with AAPM TG43 dose calculation formalism in cervical intraductal high-dose-rate brachytherapy using three different applicators
Abstract:Purpose: To compare the dosimetric effects of American Association of Physicists in Medicine (AAPM) TG43 dose formalism and AcurosBV ( grid-based Boltzmann solver ,GBBS) formalism on High-dose-rate(HDR) brachytherapy planning for cervical cancer patients irradiated using three different applicators.Methods: A TG43 plan and a AcurosBV plan were generated for each of the 30 patients. twenty patients who had undergone whole pelvic radiotherapy followed by cervical HDR brachytherapy and the remaining 10 patients w… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.