2016
DOI: 10.18869/acadpub.ijrr.14.2.149
|View full text |Cite
|
Sign up to set email alerts
|

Dosimetric comparison between XR-RV3 and EBT2 radiochromic film in megavoltage photon beams

Abstract: Background: Radiochromic film is used for radia on dose measurement, XR-RV3 is used in fluoroscopy and EBT2 film in radia on therapy. The aim was to determine if the dosimetric proper es of these two films are comparable in megavolt photon beams. Materials and Methods: Comparison measurements included: calibra on curves, heterogeneous phantom dose profiles, and nasopharynx dose distribu on measurement. Results: Both film types required 24 hours to stabilize. Their heterogeneous phantom dose profiles were simil… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
7
0

Year Published

2021
2021
2021
2021

Publication Types

Select...
3

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(7 citation statements)
references
References 13 publications
0
7
0
Order By: Relevance
“…The calibration curve performed by gradual exposure to irradiated dose, then measurement of optical density (OD) for each absolute dose, this curve was transferred to VeriSoft software and saved as a calibration curve. That will then transfer the measured map into absolute dose and map of fluence for each individual patient (18) .…”
Section: Iv-beam Fluence and Matched Map With Different Dosimeters Patient-specific Qa Ebt2 Gafchromic Filmmentioning
confidence: 99%
“…The calibration curve performed by gradual exposure to irradiated dose, then measurement of optical density (OD) for each absolute dose, this curve was transferred to VeriSoft software and saved as a calibration curve. That will then transfer the measured map into absolute dose and map of fluence for each individual patient (18) .…”
Section: Iv-beam Fluence and Matched Map With Different Dosimeters Patient-specific Qa Ebt2 Gafchromic Filmmentioning
confidence: 99%
“…These films consist of four layers: a yellow/orange polyester layer (97 µm), an adhesive layer made of acrylic (20 µm), an active layer (17 µm), and a white polyester layer (97 µm). 19 The layer sensitive (active) to the ionizing radiation is lithium salt of pentacosa-10,12-diynoic acid (LiPDCA). 19 XR-RV3 Gafchromic films are tissue-equivalent dosimetric systems, because the effective Z of all of these layers is approximately 7.3.…”
Section: Xr-rv3 Gafchromic Filmsmentioning
confidence: 99%
“…19 The layer sensitive (active) to the ionizing radiation is lithium salt of pentacosa-10,12-diynoic acid (LiPDCA). 19 XR-RV3 Gafchromic films are tissue-equivalent dosimetric systems, because the effective Z of all of these layers is approximately 7.3. 20 As a result of ionizing radiation interaction, free radicals are formed, which undergo stabilizing polymerization at least 24 hours after exposure.…”
Section: Xr-rv3 Gafchromic Filmsmentioning
confidence: 99%
See 2 more Smart Citations