Dosimetric commissioning of a high-resolution CMOS 2D detector array for patient-specific QA of single-isocenter multi-target VMAT stereotactic radiosurgery
Ching-Ling Teng,
Shih-Chi Lin,
Dale Michael Lovelock
et al.
Abstract:Stereotactic radiosurgery (SRS) using the single-isocenter-multiple-target (SIMT) technique by volumetric modulated arc therapy is increasingly popular for treating multiple brain metastases. However, the complex nature of SIMT SRS necessitates rigorous patient-specific quality assurance (PSQA). This study presents a multi-institutional dosimetric commissioning of a high-resolution complementary metal oxide semiconductor (CMOS) 2D detector array, the myQA SRS device for SIMT SRS PSQA. Basic dosimetric properti… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.