2014
DOI: 10.1117/12.2065235
|View full text |Cite
|
Sign up to set email alerts
|

Does your SEM really tell the truth? How would you know? part 3: vibration and drift

Abstract: This is the third of a series of papers discussing various causes of measurement uncertainty in scanned particle beam instruments, and some of the solutions researched and developed at NIST. Scanned particle beam instruments especially the scanning electron microscope (SEM) have gone through tremendous evolution to become indispensable tools for many and diverse scientifi c and industrial applications. These improvements have signifi cantly enhanced their performance and made them far easier to operate. But, e… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
2
0

Year Published

2016
2016
2023
2023

Publication Types

Select...
2

Relationship

0
2

Authors

Journals

citations
Cited by 2 publications
(2 citation statements)
references
References 21 publications
(21 reference statements)
0
2
0
Order By: Relevance
“…(Please refer to the web version for the interpretation of the references to color). caused by uncalibrated microscope, the drift of the stage during an acquisition, electron beam-induced contamination, and others Postek et al, 2014;Postek and Vladár, 2015) is another directly related follow-up work that also needs to be addressed. These pitfalls present a much more challenging problem, and are also commonly encountered in realworld scenarios.…”
Section: Discussionmentioning
confidence: 99%
“…(Please refer to the web version for the interpretation of the references to color). caused by uncalibrated microscope, the drift of the stage during an acquisition, electron beam-induced contamination, and others Postek et al, 2014;Postek and Vladár, 2015) is another directly related follow-up work that also needs to be addressed. These pitfalls present a much more challenging problem, and are also commonly encountered in realworld scenarios.…”
Section: Discussionmentioning
confidence: 99%
“…Installation environment factors including stray electromagnetic fields, vibration, acoustic noise, and variations in room temperature and chilling water temperature (if the SEM is water chilled), can have large effects on practical performance (Postek et al ., ). In reality, many SEMs are not housed in the buildings specifically designed to minimize the influences from the above environmental factors.…”
Section: What Sem Functions Are Needed To Understand Bulk Inorganic Mmentioning
confidence: 97%