In this paper, we introduce how to change the reaction rate as mol concentration when we scan enhanced MRI with GBCA(Gadolinium Based Contrast Agent), Also show the changing patterns depending on diverse MRI sequences which are made by different physical principle. For this study, we made MRI phantom ourselves. We mixed 500 mmol Gadoteridol with Saline in each 28 different containers from 500 to 0 mmol. After that, MR phantom was scanned by physically different MRI sequences which are T1 SE, T2 FLAIR, T1 FLAIR, 3D FLASH, T1 3D SPACE and 3D SPCIR in 1.5T bore. The results were as follows : *T1 Spin echo's Total SI(Signal Intensity) was 15608.7, Max peak was 1352.6 in 1 mmol. *T2 FLAIR's Total SI was 9106.4, Max peak was 0.4 1721.6 in 1 mmol. *T1 FLAIR's Total SI was 20972.5, Max peak was 1604.9 in 1 mmol. *3D FLASH's Total SI was 20924.0, Max peak was 1425.7 in 40 mmol. *3D SPACE 1mm's Total SI was 6399.0, Max peak was 528.3 in 3 mmol. *3D SPACE 5mm's Total SI was 6276.5, Max peak was 514.6 in 2 mmol. *3D SPCIR's Total SI was 1778.8, Max peak was 383.8 in 0.4 mmol. In most sequences, High signal intensity was shown in diluted lower concentration rather than high concentration, And also graph's max peak and pattern had difference value according to the each different sequence. Through this paper which have quantitative result of GBCA's reaction rate depending on sequence, We expect that practical enhanced MR protocol can be performed in clinical field.