SummaryWe investigated which haemodynamic parameters derived from Nexfin non-invasive continuous arterial blood pressure measurements are optimal to detect controlled volume loss in spontaneously breathing subjects. Haemodynamic monitoring was performed in 40 whole-blood donors. Mean arterial pressure, cardiac index, systemic vascular resistance index and pulse pressure variation were recorded during controlled breathing, and a Valsalva manoeuvre was performed before and after blood donation. Blood donation resulted in a reduction in cardiac index (from 3.96 AE 0.84 l.min .m 2 ; p < 0.001) and an increase in pulse pressure variation (from 13.4 AE 5.1 to 15.3 AE 5.4%; p = 0.02). The area under the receiver operating characteristic curve to detect volume loss was highest for cardiac index (0.94, 95% CI 0.88-0.99) and systemic vascular resistance (0.90, 95% CI 0.82-0.99). Nexfin is a non-invasive haemodynamic monitor that can feasibly detect volaemic changes in spontaneously breathing subjects.