Direct and fast measurement of CT beam filter profiles with simultaneous geometrical calibration

Abstract: Purpose To accurately measure the beam filter profiles from a variety of CT scanner models and to provide reference data for Monte Carlo simulations of CT scanners. Methods This study proposed a new method to measure CT beam filter profiles using a linear-array x-ray detector (X-Scan 0.8f3–512, Detection Technology Inc, Finland) under gantry rotation mode. A robust geometrical calibration approach was developed to determine key geometrical parameters, by considering the x-ray focal spot location relative to … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
10
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
6
1

Relationship

0
7

Authors

Journals

citations
Cited by 14 publications
(10 citation statements)
references
References 22 publications
(53 reference statements)
0
10
0
Order By: Relevance
“…For instance, if the TF(x) was too high, the SLRI associated with Ω x would also be too high and this can produce a value of TF(x + 1) that is too low, as indicated by Eq. (18). The resulted TFs of these two points (jump and drop) may cancel out in the next point, TF (x + 2).…”
Section: Discussionmentioning
confidence: 98%
See 2 more Smart Citations
“…For instance, if the TF(x) was too high, the SLRI associated with Ω x would also be too high and this can produce a value of TF(x + 1) that is too low, as indicated by Eq. (18). The resulted TFs of these two points (jump and drop) may cancel out in the next point, TF (x + 2).…”
Section: Discussionmentioning
confidence: 98%
“…The resulted TF will be affected even if the relative output changes by a few percent [see Eq. (18)]. This effect becomes amplified as the deviation from 360°tube rotation increases.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…An alternative approach to improving the result of tomographic reconstruction is the so-called online calibration. Methods of online calibration allow to evaluate and correct geometric errors after or directly during the tomographic experiment [16,17,18,19,20]. It can be seen that there is a gradual transition from methods with an a priori known position and shape of the calibration object (for example, [18], 2014) to methods with the simultaneous reconstruction and and geometric error correction ( [17], 2017).…”
Section: Related Workmentioning
confidence: 99%
“…In addition, Yang et al proposed a new method to measure CT filter profiles using a linear array solid-state detector for rotating x-ray sources. 35 This approach utilizes both accurate geometric calibration and high angular resolution.…”
Section: Introductionmentioning
confidence: 99%