2007
DOI: 10.1016/j.carres.2006.11.021
|View full text |Cite
|
Sign up to set email alerts
|

Differentiation of the anomeric configuration and ring form of glucosyl-glycolaldehyde anions in the gas phase by mass spectrometry: isomeric discrimination between m/z 221 anions derived from disaccharides and chemical synthesis of m/z 221 standards

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

4
50
0

Year Published

2007
2007
2019
2019

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 32 publications
(54 citation statements)
references
References 40 publications
4
50
0
Order By: Relevance
“…Given that the CID patterns of m/z 221 ions will be used for structural identification, spectral reproducibility is an important issue. Similar to the findings from a Paul trap instrument [33,34], we noticed that the number of ions (m/z 221) in the linear ion trap and the energy input into an ion were among the most important parameters affecting spectral reproducibility. To ensure reasonable ion statistics and avoid adverse space charge effects, the intensity of the m/z 221 ions was kept at 1× 10 6 cps before MS 3 CID.…”
Section: Cid Of M/z 221 Ions Generated From 1-3 Linked Disaccharidessupporting
confidence: 80%
See 4 more Smart Citations
“…Given that the CID patterns of m/z 221 ions will be used for structural identification, spectral reproducibility is an important issue. Similar to the findings from a Paul trap instrument [33,34], we noticed that the number of ions (m/z 221) in the linear ion trap and the energy input into an ion were among the most important parameters affecting spectral reproducibility. To ensure reasonable ion statistics and avoid adverse space charge effects, the intensity of the m/z 221 ions was kept at 1× 10 6 cps before MS 3 CID.…”
Section: Cid Of M/z 221 Ions Generated From 1-3 Linked Disaccharidessupporting
confidence: 80%
“…Ion trap CID of deprotonated 1-3 linked disaccharides (m/z 341) typically generates ions at m/z 221 in trace abundance on a Paul trap instrument, and isolation or further CID of m/z 221 ions have not been achieved before [14,15,34,36]. The 4000QTRAP mass spectrometer used in this study has a unique triple quadrupole-linear ion trap configuration, offering high sensitivity due to the large capacity of the linear ion trap, and allowing either beam-type or ion trap collisional activation.…”
Section: Resultsmentioning
confidence: 99%
See 3 more Smart Citations