volume 55, issue 4, P256-259 2011
DOI: 10.1590/s0004-27302011000400003
View full text
|
|
Share

Abstract: OBJECTIVE: To evaluate the accuracy of potassium concentrations measured by blood gas analysis (PBG) compared with laboratory serum potassium (LSP), in the initial care of patients with diabetic ketoacidosis (DKA). SUBJECTS AND METHODS: Fifty three patients with diabetes mellitus were evaluated in a retrospective analysis. PBG was carried out using the Radiometer ABL 700 (Radiometer Copenhagen®), and results were compared with LSP ADVIA 1650 Chemistry system (Siemens®), the gold standard method. Both methods a…