“…4 Each pixel in the array was able to resolve photon energies to a resolution of better than 5 eV. This resolution was achieved in part by operating the array at a very low temperature, 50 mK, using a three-stage adiabatic demagnetization refrigerator (ADR) whose operation was supported by a cryogenic system 5 consisting of a 40-L superfluid helium tank (at <1.3 K) and a 4.5 K Joule-Thomson (JT) cryocooler, with additional two-stage Stirling cryocoolers 6,7 used to cool vapor-cooled-shields and provide precooling for the JT stage.…”
Section: Introductionmentioning
confidence: 99%
“…4 Each pixel in the array was able to resolve photon energies to a resolution of better than 5 eV. This resolution was achieved in part by operating the array at a very low temperature, 50 mK, using a three-stage adiabatic demagnetization refrigerator (ADR) whose operation was supported by a cryogenic system 5 consisting of a 40-L superfluid helium tank (at <1.3 K) and a 4.5 K Joule-Thomson (JT) cryocooler, with additional two-stage Stirling cryocoolers 6,7 used to cool vapor-cooled-shields and provide precooling for the JT stage.…”
Section: Introductionmentioning
confidence: 99%
“…Figure 2 shows this cooling system and its complex arrangement of both active and passive cooling elements, including four 20-K 2-stage Stirling cryocoolers (2ST), 19 a 4.5-K Joule-Thomson (JT) cryocooler, 19 a 40-L superfluid helium tank, and 3-stage adiabatic demagnetization refrigerator (ADR). 9 When integrated, these elements are nested within a multistage dewar featuring four helium vapor cooled shields: outer, middle, and inner (OVCS, MVCS, and IVCS), and the innermost JT shield surrounding the helium tank.…”
Section: Cooling System Overviewmentioning
confidence: 99%
“…In addition, the spectrum of the 2ST compressor contains a continuum feature above ∼200 Hz, caused by frictional motion, of the level of ∼2 N rms integrated in 0 to 500 Hz for each axis. 11 The vibrational heating occurs because of the coincidence of vibration spectrum and the sensitive frequency.…”
Section: Mechanism Of How Energy Resolution Was Degradedmentioning
confidence: 99%
“…Both the 2STs and the JT are mounted on the outer surface of DMS. 7,11 When the cryocoolers are in operation, the compressors and cold heads warm up because of the electrical power applied. The dissipated heat is transported to either the dewar surface or radiator panels and radiated to deep space.…”
Section: Overview Of the Sxs Cooling Systemmentioning
confidence: 99%