2017
DOI: 10.1016/j.jviromet.2017.05.002
|View full text |Cite
|
Sign up to set email alerts
|

Development of the Abbott RealTime ZIKA assay for the qualitative detection of Zika virus RNA from serum, plasma, urine, and whole blood specimens using the m2000 system

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
9
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 13 publications
(9 citation statements)
references
References 34 publications
0
9
0
Order By: Relevance
“…The final ZIKV rRT-PCR assay with current peer-reviewed literature is the Abbott RealTime ZIKV RT-PCR, which targets the prM and NS3 genes and can be performed with WB, plasma, serum, and urine samples, using the mSample RNA preparation system kit and automated m2000sp and m2000rt instruments (Abbott Molecular Inc., Des Planes IL). Compared to the Altona RealStar ZIKV rRT-PCR, the Abbott ZIKV rRT-PCR demonstrated positive and negative agreement rates of 96.5% and 71.4% in serum samples and 72.4% and 71.4% in urine samples, respectively, with LLoD values of 30 copies/ml for serum samples and 40 copies/ml for urine and plasma samples, as reflected in the assay IFU (58).…”
Section: Performance Of Molecular Assays With Fda Eua For Detection Omentioning
confidence: 93%
“…The final ZIKV rRT-PCR assay with current peer-reviewed literature is the Abbott RealTime ZIKV RT-PCR, which targets the prM and NS3 genes and can be performed with WB, plasma, serum, and urine samples, using the mSample RNA preparation system kit and automated m2000sp and m2000rt instruments (Abbott Molecular Inc., Des Planes IL). Compared to the Altona RealStar ZIKV rRT-PCR, the Abbott ZIKV rRT-PCR demonstrated positive and negative agreement rates of 96.5% and 71.4% in serum samples and 72.4% and 71.4% in urine samples, respectively, with LLoD values of 30 copies/ml for serum samples and 40 copies/ml for urine and plasma samples, as reflected in the assay IFU (58).…”
Section: Performance Of Molecular Assays With Fda Eua For Detection Omentioning
confidence: 93%
“…However, as this method allows the simultaneous detection of both specific and nonspecific PCR products, it produces false positives. Frankel et al (2017) developed the Abbott Real-Time ZIKA assay, and the detection limit of this assay was shown to be 120 copy/lL in whole blood. Moreover, Balm et al (2012) developed a sensitive and specific one-step RT-qPCR assay, and the lower limit of detection of this assay was 140 RNA copy/PCR.…”
Section: Discussionmentioning
confidence: 99%
“…The assay was highly specific and sensitive with a limit of detection (LOD) of 120 copies/ml in whole blood, 40 copies/ml in urine and plasma, and 30 copies/ml in serum 399 . While, Hancock et al 400 used electronic health records (RT‐PCR) in a syndromic surveillance program to monitor asymptomatic pregnant women from American Samoa in order to establish a timeline to cease routine testing.…”
Section: Molecular Assaysmentioning
confidence: 99%