Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials

Abstract: An anthropomorphic head phantom, constructed from a water‐equivalent plastic shell with only a spherical target, was modified to include a nonspherical target (pituitary) and an adjacent organ at risk (OAR) (optic chiasm), within 2 mm, simulating the anatomy encountered when treating acromegaly. The target and OAR spatial proximity provided a more realistic treatment planning and dose delivery exercise. A separate dosimetry insert contained two TLD for absolute dosimetry and radiochromic film, in the sagittal … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

2
19
0

Year Published

2015
2015
2024
2024

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 13 publications
(21 citation statements)
references
References 16 publications
2
19
0
Order By: Relevance
“…12 The results presented in this study also demonstrate similar levels of agreement with the TPS to other studies that also used anthropomorphic phantoms. 8,16,27 Most importantly, the film measurements show a high degree of consistency which is essential if the methodology is to be employed for audit purposes. The high agreement with the TPS and consistency in measurement suggests that this is an appropriate phantom design and film methodology to be used for an end-to-end audit.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…12 The results presented in this study also demonstrate similar levels of agreement with the TPS to other studies that also used anthropomorphic phantoms. 8,16,27 Most importantly, the film measurements show a high degree of consistency which is essential if the methodology is to be employed for audit purposes. The high agreement with the TPS and consistency in measurement suggests that this is an appropriate phantom design and film methodology to be used for an end-to-end audit.…”
Section: Discussionmentioning
confidence: 99%
“…6 There are numerous examples where anthropomorphic phantoms have been utilized for complex radiotherapy treatment quality assurance, audits and trials. [7][8][9][10] In this study, we present the adaptation of a commercial anthropomorphic phantom for the novel use of three simultaneous detectors, with the purpose of employing it for a radiosurgery dosimetry audit. The methodology utilized was guided by the results of a recent survey on the current practice for cranial SRS in the UK.…”
Section: Introductionmentioning
confidence: 99%
“…Stereotactic radiosurgery (SRS), by definition, is the delivery of a large therapeutic dose of radiation to an accurately localized target 1 , 2 , 3 , 4 , 5 , 6 , 7 , 8 . This results in an ablative dose of radiation at the target with a steep dose gradient to minimize damage to tissues outside of the target 9 , 10 , 11 , 12 .…”
Section: Introductionmentioning
confidence: 99%
“…The Leksell Gamma Knife consists of 201 collimated cobalt 60 sources arranged over a hemispherical surface . To implement Gamma Knife radiosurgery in clinical practice, some quality control and dosimetry should be performed . Several studies were performed to measure the dose delivered to the target with a homogenous head phantom and some dosimeters such as radichromic film, ion chamber, diode, etc .…”
Section: Introductionmentioning
confidence: 99%
“…3 Several studies were performed to measure the dose delivered to the target with a homogenous head phantom and some dosimeters such as radichromic film, ion chamber, diode, etc. [1][2][3] These studies used a modified head phantom similar to an adult human head with the actual size of a head. These phantoms were made from water or water-equivalent soft tissue, and no heterogeneity was considered.…”
Section: Introductionmentioning
confidence: 99%