2013
DOI: 10.1016/j.apradiso.2013.03.032
|View full text |Cite
|
Sign up to set email alerts
|

Development of a low-level radon reference atmosphere

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
11
0

Year Published

2014
2014
2023
2023

Publication Types

Select...
6
1

Relationship

2
5

Authors

Journals

citations
Cited by 10 publications
(11 citation statements)
references
References 3 publications
0
11
0
Order By: Relevance
“…Figure 1 shows the raw-counts that were computed from the spectra (input data), the relative humidity inside the chamber as measured by a SHT-35 sensor (red curve), the inferred filtering distributions and the results that would have been obtained from the method in [1,2]. It can be observed from these results that once the dynamics of ingrowth are modeled in this way it becomes apparent that the methods in [1,2] lead to deviations from the true value. The new method extends the validity to regimes of nonconstant 𝜂𝜂 and allows for an estimate of the naturally expected increased uncertainty in these regimes.…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…Figure 1 shows the raw-counts that were computed from the spectra (input data), the relative humidity inside the chamber as measured by a SHT-35 sensor (red curve), the inferred filtering distributions and the results that would have been obtained from the method in [1,2]. It can be observed from these results that once the dynamics of ingrowth are modeled in this way it becomes apparent that the methods in [1,2] lead to deviations from the true value. The new method extends the validity to regimes of nonconstant 𝜂𝜂 and allows for an estimate of the naturally expected increased uncertainty in these regimes.…”
Section: Resultsmentioning
confidence: 99%
“…Emanation sources are 226 Ra sources constructed so that some fraction of the generated 222 Rn is released from them. In [1,2], an approach to measure the released amount of 222 Rn based on measuring the residual 222 Rn in the source is presented. However, this approach is only valid for times in which steady state has been reached.…”
Section: Introductionmentioning
confidence: 99%
“…Thus, the components of the radon reference chamber are made of stainless steel. Therefore, there is no chance for any additional source of radon to enter (Linzmaier, and Röttger 2013). For this procedure, if the radon gas emanation is constant, the emanation factor χ is calculated using: (1)…”
Section: Experimental Set-up and Proceduresmentioning
confidence: 99%
“…To reduce the measurement uncertainty and the limited statistics of the count rate, a low level radon reference chamber with specific measurement equipment was set up (Linzmaier, 2012 andLinzmaier andRöttger, 2013). Therefore, a large volume detector was developed as a transfer standard for the realization of the unit Bq/m 3 .…”
Section: Introductionmentioning
confidence: 99%