Development of a full-ring “add-on PET” prototype: A head coil with DOI-PET detectors for integrated PET/MRI

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

1
23
0

Year Published

2019
2019
2023
2023

Publication Types

Select...
4
3

Relationship

0
7

Authors

Journals

citations
Cited by 24 publications
(24 citation statements)
references
References 21 publications
1
23
0
Order By: Relevance
“…Subsequently, the LYSO crystal dimensions were reduced to 2 • 2 • 5 mm 3 , and they were arranged in 4 layers of 19 • 6 arrays ( Fig. 5C) (57). The PET detector modules were mounted between the elements of a 3-T MRI birdcage coil to evaluate their performance.…”
Section: Conventional Scanner Geometry Devicesmentioning
confidence: 99%
See 1 more Smart Citation
“…Subsequently, the LYSO crystal dimensions were reduced to 2 • 2 • 5 mm 3 , and they were arranged in 4 layers of 19 • 6 arrays ( Fig. 5C) (57). The PET detector modules were mounted between the elements of a 3-T MRI birdcage coil to evaluate their performance.…”
Section: Conventional Scanner Geometry Devicesmentioning
confidence: 99%
“…The spatial resolution ranged from 2.3 mm at the center of the FOV to 3.5 mm offcenter. The sensitivity was not estimated given the relatively limited extent of the axial FOV (1.2 cm), which would not be sufficient for actual brain studies, as acknowledged by the authors (57). A different approach to minimizing the interference between the two modalities was proposed by researchers from Stanford.…”
Section: Conventional Scanner Geometry Devicesmentioning
confidence: 99%
“…7,15 Some studies have recently reported on PET/MRS imaging as well. [16][17][18] Although the whole-body PET/MRI system became available in 2010, 9 recent growing interest in the development of a low-cost PET insert (such as a brain PET) for existing MRI systems [19][20][21][22][23][24] has created a potentially dominant new research field in multimodal medical imaging research. However, in all these PET insert developments, a singlechannel volume radiofrequency (RF) coil (e.g., birdcage RF coil) was implemented as a transmitter.…”
Section: Major Papermentioning
confidence: 99%
“…[25][26][27][28][29][30][31] Conventionally, detectors and front-end electronics of a PET insert are RF shielded to avoid electromagnetic (EM) interference between the two systems. [19][20][21][22][23][24] In this study, we developed a novel design of a microstrip transmission line RF coil [27][28][29][30] that combines the RF shield of a PET detector module with a microstrip conductor that acts as a transmission line coil. The conventional microstrip coil consists of two metal conductors (one is a microstrip conductor, and the other is a ground conductor) and a dielectric between the two conductors.…”
Section: Major Papermentioning
confidence: 99%