2017
DOI: 10.1007/s10544-017-0218-8
|View full text |Cite
|
Sign up to set email alerts
|

Development of a biodegradable flow resisting polymer membrane for a novel glaucoma microstent

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
12
0
1

Year Published

2018
2018
2023
2023

Publication Types

Select...
7
1

Relationship

2
6

Authors

Journals

citations
Cited by 10 publications
(13 citation statements)
references
References 36 publications
0
12
0
1
Order By: Relevance
“…Initial flow resistance of the resorbable nonwoven (1.543 mmHg/µl min -1 ) conforms well with the target value of 1.850 mmHg/μl min -1 corresponding to a physiological aqueous humour flow rate of 2 µl min -1 and a physiological pressure difference between the anterior chamber of the eye and the suprachoroidal space of 3.7 mmHg [6]. Therefore, the risk of postoperative hypotony is minimized [5]. Chronosil 80A as base material of the drugeluting coating allows for a well-controlled drug release.…”
Section: Discussionmentioning
confidence: 56%
See 2 more Smart Citations
“…Initial flow resistance of the resorbable nonwoven (1.543 mmHg/µl min -1 ) conforms well with the target value of 1.850 mmHg/μl min -1 corresponding to a physiological aqueous humour flow rate of 2 µl min -1 and a physiological pressure difference between the anterior chamber of the eye and the suprachoroidal space of 3.7 mmHg [6]. Therefore, the risk of postoperative hypotony is minimized [5]. Chronosil 80A as base material of the drugeluting coating allows for a well-controlled drug release.…”
Section: Discussionmentioning
confidence: 56%
“…Resorbable nonwovens were manufactured from Poly(4hydroxybutyrate) (P(4HB); P4HB biopolymer, Tepha, Inc., Lexington, MA, USA) by means of electrospinning as described previously [5]. Process parameters of electrospinning are summarized in Table 1.…”
Section: Resorbable Nonwovenmentioning
confidence: 99%
See 1 more Smart Citation
“…В качестве мер борьбы с указанными недостатка-ми применяется насыщение дренажей лекарственными препаратами [25], модификация формы [26] и поверхно-сти дренажей [20] или переход к резорбируемой основе [27]. Находясь в интрасклеральной полости в заданный временной интервал, резорбируемые дренажи способ-ствуют более совершенному формированию путей от-тока, тем самым пролонгируя гипотензивный эффект операции.…”
Section: Discussionunclassified
“…Glaucoma is a major eye disorder causing irreversible blindness treated by important device inventions in industrialized countries [1][2][3]. In this research, basic concepts of adaptive implant for minimally invasive glaucoma treatment [4] were studied as shown in Fig.…”
Section: Introductionmentioning
confidence: 99%