1997
DOI: 10.1155/s1463924697000187
|View full text |Cite
|
Sign up to set email alerts
|

Development and analytical performance of an automated screening method for cannabinoids on the Dimension clinical chemistry system

Abstract: A fully automated, random access method for the determination of cannabinoids (UTHC) was developed for the Dimension AR and XL clinical chemistry systems. The method utilizes Abuscreen ONLINE reagents and a multianalyte liquid calibrator containing 11-nor-Δ9-THC-9-carboxylic acid. Within-run and total reproducibility, determined using NCCLS protocol EP5- T2, was less than 0.6% and 1.6% CV, respectively, at all concentrations. Calibration stability was retained for at least 30 days. An extensive evaluation of n… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...

Citation Types

0
0
0

Year Published

2000
2000
2000
2000

Publication Types

Select...
1

Relationship

0
1

Authors

Journals

citations
Cited by 1 publication
references
References 7 publications
(7 reference statements)
0
0
0
Order By: Relevance