Abstract:Tube-based-polarized energy-dispersive X-ray fluorescence (EDXRF) is a powerful adaptation on traditional EDXRF, requiring very specific geometry and a scattering target to generate polarized X-rays. This secondary target is typically chosen to be a metallic foil, allowing for the polarization of the incident X-ray beam, and the addition of the secondary target's fluorescence response to the initial beam. A simulation, using GEANT4 Monte Carlo code, and an experimental confirmation were used to determine the o… Show more
“…Large differences can be seen in the K, Fe, and Zn peaks, with small apparent differences in the Ca and Cu peaks. The Ni and Ti peaks come from contaminants in the measuring system…”
Section: Resultsmentioning
confidence: 99%
“…The XRF measurements were taken using a PEDXRF system developed and optimized specifically for measuring soft biological tissues . PEDXRF systems allow for high fluence rate, multielemental analysis of samples .…”
Section: Methodsmentioning
confidence: 99%
“…The XRF measurements were taken using a PEDXRF system developed and optimized specifically for measuring soft biological tissues . PEDXRF systems allow for high fluence rate, multielemental analysis of samples . The newest generation of PEDXRF spectrometers allows for highly accurate results to be measured on benchtop systems in reasonable time frames, removing the need of a synchrotron radiation source to perform these XRF studies .…”
Section: Methodsmentioning
confidence: 99%
“…PEDXRF systems allow for high fluence rate, multielemental analysis of samples . The newest generation of PEDXRF spectrometers allows for highly accurate results to be measured on benchtop systems in reasonable time frames, removing the need of a synchrotron radiation source to perform these XRF studies . Similar to synchrotron sources, PEDXRF systems use a polarized X‐ray beam to excite the sample, reducing scatter incident on the detector, allowing for higher power sources to be utilized.…”
Section: Methodsmentioning
confidence: 99%
“…Similar to synchrotron sources, PEDXRF systems use a polarized X‐ray beam to excite the sample, reducing scatter incident on the detector, allowing for higher power sources to be utilized. The secondary target, which produces the polarized X‐ray source, can also be chosen to optimally excite elements of interest in the sample …”
Previous studies have shown that the combination of X‐ray fluorescence and X‐ray diffraction data can be used as a histopathological characterization tool for breast tissue. Recent advances in energy‐dispersive X‐ray fluorescence techniques have allowed for benchtop systems to produce useful results in a reasonable time frame, allowing for clinical implementation to be realized. Using a polarized energy‐dispersive X‐ray fluorescence and energy‐dispersive X‐ray diffraction system optimized for measuring soft tissues, 38 breast tissue samples (19 normal and 19 diseased) were interrogated. The measured elemental concentrations and adipose and fibrous tissue contents were used in a principal component analysis study to determine the variables that produced the most differentiation between the normal and diseased tissues. For each sample, a soft independent modeling of class analogy technique was utilized to create classification models using the K, Fe, and Zn concentration and adipose and fibrous tissue content of all other breast samples. The class model produced from both X‐ray fluorescence and X‐ray diffraction data correctly classified 31 of 38 samples with no false positives or false negatives, showing improvement from solely X‐ray fluorescence models or X‐ray diffraction models alone, and demonstrates the usefulness of such a technique.
“…Large differences can be seen in the K, Fe, and Zn peaks, with small apparent differences in the Ca and Cu peaks. The Ni and Ti peaks come from contaminants in the measuring system…”
Section: Resultsmentioning
confidence: 99%
“…The XRF measurements were taken using a PEDXRF system developed and optimized specifically for measuring soft biological tissues . PEDXRF systems allow for high fluence rate, multielemental analysis of samples .…”
Section: Methodsmentioning
confidence: 99%
“…The XRF measurements were taken using a PEDXRF system developed and optimized specifically for measuring soft biological tissues . PEDXRF systems allow for high fluence rate, multielemental analysis of samples . The newest generation of PEDXRF spectrometers allows for highly accurate results to be measured on benchtop systems in reasonable time frames, removing the need of a synchrotron radiation source to perform these XRF studies .…”
Section: Methodsmentioning
confidence: 99%
“…PEDXRF systems allow for high fluence rate, multielemental analysis of samples . The newest generation of PEDXRF spectrometers allows for highly accurate results to be measured on benchtop systems in reasonable time frames, removing the need of a synchrotron radiation source to perform these XRF studies . Similar to synchrotron sources, PEDXRF systems use a polarized X‐ray beam to excite the sample, reducing scatter incident on the detector, allowing for higher power sources to be utilized.…”
Section: Methodsmentioning
confidence: 99%
“…Similar to synchrotron sources, PEDXRF systems use a polarized X‐ray beam to excite the sample, reducing scatter incident on the detector, allowing for higher power sources to be utilized. The secondary target, which produces the polarized X‐ray source, can also be chosen to optimally excite elements of interest in the sample …”
Previous studies have shown that the combination of X‐ray fluorescence and X‐ray diffraction data can be used as a histopathological characterization tool for breast tissue. Recent advances in energy‐dispersive X‐ray fluorescence techniques have allowed for benchtop systems to produce useful results in a reasonable time frame, allowing for clinical implementation to be realized. Using a polarized energy‐dispersive X‐ray fluorescence and energy‐dispersive X‐ray diffraction system optimized for measuring soft tissues, 38 breast tissue samples (19 normal and 19 diseased) were interrogated. The measured elemental concentrations and adipose and fibrous tissue contents were used in a principal component analysis study to determine the variables that produced the most differentiation between the normal and diseased tissues. For each sample, a soft independent modeling of class analogy technique was utilized to create classification models using the K, Fe, and Zn concentration and adipose and fibrous tissue content of all other breast samples. The class model produced from both X‐ray fluorescence and X‐ray diffraction data correctly classified 31 of 38 samples with no false positives or false negatives, showing improvement from solely X‐ray fluorescence models or X‐ray diffraction models alone, and demonstrates the usefulness of such a technique.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.