Determination of Full Energy Peak Efficiency of NaI(Tl) Detector Depending on Efficiency Transfer Principle for Conversion Form Experimental Values

Abstract: In this work we calibrated the NaI(Tl) scintillation detectors (5.08 × 5.08 cm<sup>2</sup> and 7.62 × 7.62 cm<sup>2</sup>) and the Full Energy Peak Efficiency (FEPE) for these detectors have been calculated for point sources placed at different positions on the detector axis using the analytical approach of the effective solid angle ratio. This approach is based on the direct mathematical method reported by Selim and Abbas [1,2] and has been used successfully before to calibrate the cyl… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
11
0

Year Published

2013
2013
2024
2024

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 25 publications
(11 citation statements)
references
References 9 publications
0
11
0
Order By: Relevance
“…The signals from the detector were amplified and analyzed with an 8K multichannel analyzer. The efficiency of the NaI (Tl) detector is higher at low source energy (26). The Low-Z samples such as sodium thiosulfate (Na 2 S 2 O 3 ), benzoic acid (C 7 H 6 O 2 ), sodium hydroxide (NaOH), poly vinyl alcohol (PVA) (C 2 H 4 O), potassium nitrate(KNO3), naphthalene (C 10 H 8 ) using KBr press machine pellets were prepared having the same thickness.…”
Section: Methodsmentioning
confidence: 99%
“…The signals from the detector were amplified and analyzed with an 8K multichannel analyzer. The efficiency of the NaI (Tl) detector is higher at low source energy (26). The Low-Z samples such as sodium thiosulfate (Na 2 S 2 O 3 ), benzoic acid (C 7 H 6 O 2 ), sodium hydroxide (NaOH), poly vinyl alcohol (PVA) (C 2 H 4 O), potassium nitrate(KNO3), naphthalene (C 10 H 8 ) using KBr press machine pellets were prepared having the same thickness.…”
Section: Methodsmentioning
confidence: 99%
“…To reduce the background radiation, the inner parts of the detector were surrounded by Pb shielding, as shown in Figure 1(b). This geometry was adapted from Abd-Elzaher, et al [25].…”
Section: Methodsmentioning
confidence: 99%
“…This is likely because the NaI-efficiency decreases smoothly with energy for energies greater than 120 keV. 30 If additional data points were measured at energies below 120 keV, a function with more than two parameters might be necessary to accurately describe the efficiency of the thyroid-probe.…”
Section: A1 Thyroid-probe Efficiency Curvementioning
confidence: 99%