Detection of Electrode Proximity to the Cochlea Wall Based on Impedance Variation: a Preliminary Computational Study

Abstract: The electro anatomy of the cochlea plays a crucial role in hearing, where damage to the cochlea may cause hearing loss. Cochlear implants provide hearing to severe or profound hearing-impaired individuals. The accurate insertion of electrodes into the cochlea is important. If misplaced it may lead to further damage (insertion trauma). Visual inspection of the electrode array insertion is limited and relies on the experience of the surgeon. Assisted real time guidance in positioning the electrode array in the c… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2

Citation Types

0
5
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
5
1

Relationship

3
3

Authors

Journals

citations
Cited by 6 publications
(7 citation statements)
references
References 12 publications
0
5
0
Order By: Relevance
“…Model 2 and Model 3 evaluate the centerline tracing algorithm against the "golden standard," i.e., a hand-traced centerline by clinicians in realistic reconstructed cochlea models. The realistic cochlea models were reconstructed using micro-CT images of 512 × 512 pixels per slice databases from the Royal National Throat, Nose and Ear Hospital, and a manually defined ground-truth was used to quantify traversal performance [39]. [31], Active contour [32]- [35] and Supercylinder.…”
Section: Results and Validationmentioning
confidence: 99%
See 1 more Smart Citation
“…Model 2 and Model 3 evaluate the centerline tracing algorithm against the "golden standard," i.e., a hand-traced centerline by clinicians in realistic reconstructed cochlea models. The realistic cochlea models were reconstructed using micro-CT images of 512 × 512 pixels per slice databases from the Royal National Throat, Nose and Ear Hospital, and a manually defined ground-truth was used to quantify traversal performance [39]. [31], Active contour [32]- [35] and Supercylinder.…”
Section: Results and Validationmentioning
confidence: 99%
“…In each tracing trial, centerline coordinates and width deviations are recorded and compared with the ground truth. The traced centerline coordinates present the optimal path for surgical purposes as studied in [39]- [40], where an impedance model estimates the proximity of an inserted surgical electrode array to the cochlea wall.…”
Section: Results and Validationmentioning
confidence: 99%
“…A transmitter sends sound signals to a receiver and stimulator implanted under the skin. They stimulate the auditory nerve with electrodes that have been threaded into the cochlea [13] .…”
Section: Introductionmentioning
confidence: 99%
“…Also, using conventional techniques it may not be feasible to conduct systematic comparison across individuals to examine the precise position of the electrode (Pile et al, 2017 ). Computational cochlea models have been utilized to simulate the current spread in the cochlea and provided useful insights (Malherbe et al, 2016 ; Salkim et al, 2020 ). Such models are implemented using the finite element method (FEM).…”
Section: Introductionmentioning
confidence: 99%