2003
DOI: 10.1109/tns.2003.817950
|View full text |Cite
|
Sign up to set email alerts
|

Design evaluation of A-PET: A high sensitivity animal PET camera

Abstract: In recent years it has been shown that PET is capable of obtaining in vivo metabolic images of small animals. These serve as models to study the development and progress of diseases within humans. Imaging small animals requires not only image resolution better than 2 mm, but also high sensitivity in order to image ligands with low specific activity or radiochemical yields. Toward achieving these goals, we have developed a discrete 2 2 10 mm 3 GSO Anger-logic detector for use in a high resolution, high sensitiv… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
51
0
3

Year Published

2005
2005
2023
2023

Publication Types

Select...
6
3

Relationship

0
9

Authors

Journals

citations
Cited by 91 publications
(54 citation statements)
references
References 16 publications
0
51
0
3
Order By: Relevance
“…The scanner was one of two prototype, preproduction units, the performance characteristics of which have recently been described (12). In brief, the scanner has an effective axial field of view of 11.6 cm and resolution measured at 2.26 mm at the center of the field of view.…”
Section: Pet Imagingmentioning
confidence: 99%
“…The scanner was one of two prototype, preproduction units, the performance characteristics of which have recently been described (12). In brief, the scanner has an effective axial field of view of 11.6 cm and resolution measured at 2.26 mm at the center of the field of view.…”
Section: Pet Imagingmentioning
confidence: 99%
“…To monitor the changes of SPIO signal over time, MR imaging was performed in group 1 at day 1 after injection and then weekly for 4 weeks. To compare ventricular function in ESC-treated and culture mediatreated hearts, MR imaging was performed in groups 2 and 3 at day 1, at which time infarct size was also quantified, and at 4 weeks after surgery.PET imaging was performed with a high-resolution (2-mm) dedicated smallanimal PET scanner, which was based on a commercially manufactured unit (Mosaic; Philips Medical Systems, Cleveland, Ohio) (15,16 MR imaging was performed with a 4.7-T horizontal bore magnet (Magnex Scientific, Walnut Creek, Calif) interfaced with a console (Inova; Varian, Palo Alto, Calif). To assess global function and SPIO-associated hypoenhancement, a T2*-weighted fast gradientecho sequence (17) with a 20°flip angle, 3-msec echo time (to minimize cardiac motion), and effective repetition time equal to one heartbeat was * The first time for detection of 1 and 5 million cells is set differently, because on the basis of the pilot study, the earliest time to detect 5 million cells in the heart was day 4 after injection, whereas the earliest time to detect 1 million cells was day 7 after injection (Appendix E1 ͓http://radiology.rsnajnls.org/cgi/content/full/250/3/821/DC1͔).…”
mentioning
confidence: 99%
“…Calculations of the energy resolution of crystal LYSO10-P2 were performed, using separately measured values for the individual contributions in (5). The results of these calculations were compared to measured energy resolutions.…”
Section: B Laapd Energy Resolutionmentioning
confidence: 99%
“…Calculations according to (5) were also done, again using separately measured values of the individual contributions.…”
Section: ) Time Resolutionmentioning
confidence: 99%