Design and Optimization of a Linear Wavenumber Spectrometer with Cylindrical Optics for Line Scanning Optical Coherence Tomography

Abstract: We report the design of a high-efficiency spectral-domain spectrometer with cylindrical optics for line scanning optical coherence tomography (OCT). The spectral nonlinearity in k space (wavenumber) lowers the depth-dependent signal sensitivity of the spectrometers. For linearizing, in this design, grating and prism have been introduced. For line scanning, a cylindrical mirror is utilized in the scanning part. Line scanning improves the speed of imaging compared to fly-spot scanning. Line scanning OCT requires… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
2
1

Relationship

0
3

Authors

Journals

citations
Cited by 3 publications
(1 citation statement)
references
References 30 publications
(55 reference statements)
0
1
0
Order By: Relevance
“…The resolution and bandwidth of the SD-OCT system are often limited by the spectrometer. Grating spectrometers are widely used in SD-OCT [45] and its performance is determined by various physical parameters such as grating length, number of gratings, material used, wavelengths, etc. [46].…”
Section: Discussionmentioning
confidence: 99%
“…The resolution and bandwidth of the SD-OCT system are often limited by the spectrometer. Grating spectrometers are widely used in SD-OCT [45] and its performance is determined by various physical parameters such as grating length, number of gratings, material used, wavelengths, etc. [46].…”
Section: Discussionmentioning
confidence: 99%