Danger of low pressure alarm failure in preterm infants on continuous positive airway pressure

Abstract: In certain settings of conventional continuous positive airway pressure (CPAP) application, the ventilator may not be able to detect dislodgement of the prongs. This occurs especially in settings with high flow and small prongs. We investigated the relation between ventilator flows, size of the nasal prongs, and pressure generated within the ventilator circuit due to the flow resistance of the prongs. We studied a Baby-flow CPAP connected to a Babylog 8000plus ventilator. Five prongs of increasing size (x-smal… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2011
2011
2024
2024

Publication Types

Select...
3

Relationship

1
2

Authors

Journals

citations
Cited by 3 publications
(1 citation statement)
references
References 14 publications
0
1
0
Order By: Relevance
“…A further problem occurs in case of leakage within the airways of the patient. In this case, the CPAP system tries to compensate for the pressure loss by increasing the flow (22). This increased flow leads, in turn, to an incalculable increase of R E .…”
Section: Discussionmentioning
confidence: 99%
“…A further problem occurs in case of leakage within the airways of the patient. In this case, the CPAP system tries to compensate for the pressure loss by increasing the flow (22). This increased flow leads, in turn, to an incalculable increase of R E .…”
Section: Discussionmentioning
confidence: 99%