Case 1A 65-year-old patient was referred for pacemaker implantation due to sick sinus syndrome. The implantation of a DDDR pacemaker (Kappa 901, Medtronic, Minneapolis, MN, USA) was uneventful, and two straight active fixation leads were introduced and placed, one in the right atrial appendage and the other in the right ventricular outflow tract. During implantation, immediately after the leads were connected to the pacemaker, a change in heart rhythm was observed from a sinus rhythm of 75 bpm with a first degree atrioventricular (AV) block to an apparently atrial paced rhythm of 125 bpm (Fig. 1). It was soon established that the atrial lead was placed in the ventricular port (VP) and the ventricular lead was placed in the atrial port. The situation was easily and promptly resolved.
Case 2A 77-year-old man with a symptomatic second and third degree atrioventricular block underwent DDD pacemaker implantation (Kairos D, Biotronik, Berlin, Germany). The procedure was uneventful, with an apparently good, fully paced rhythm of 60 bpm at the end of the procedure. However, several hours after implantation the patient suffered a syncopal attack. The electrocardiogram (ECG) (Fig. 2) obtained at that time indicated periods of asystole with occasional ventricular capture following the atrial pacing spike. The pacemaker was reprogrammed to the AAI pacing mode and the symptoms resolved. The next day, after confirming under fluoroscopy that the leads were appropriately positioned in the heart, the pacemaker pocket was opened and, as expected, the leads were found to be incorrectly positioned (switched) in the pacemaker header. After correction there were no further sequelae and the patient remained asymptomatic during the subsequent 7 years of followup.What was the mechanism of such different presentations despite the same nature of the problem in these two cases ?
DiscussionAccidental reversal of the atrial and ventricular leads in the pacemaker header is a rare and embarrassing complication, and is therefore probably underreported. Consequently, the observed clinical manifestations are poorly documented with only two case reports published. 1,2 A similar mistake with the proximal and distal defibrillator coil terminal pins erroneously switched in the header was also reported. 3 In the first case, the initial thought was that the pacemaker was inadvertently reprogrammed from the default program (DDD, lower rate 60/min, sensed AV delay 120 ms) to either AAIR or DDDR with a long AV delay and the sensor accelerated the pacing rate due to device movement during placement in the pacemaker pocket. However, since the fast paced rhythm did not slow down in the following few minutes, an alternative explanation was sought. Both the QRS-spike and the spike-QRS intervals were constant during the arrhythmia with the QRS-spike interval being equal to the sensed AV delay. This strongly suggested that the pacing spike came from the ventricular channel and was coupled with both the preceding and the following QRS. Such a situation can be pr...