Correlation functions for Elekta aSi EPIDs used as transit dosimeter for open fields

Abstract: In‐vivo dosimetry techniques are currently being applied only by a few Centers because they require time‐consuming implementation measurements, and workload for detector positioning and data analysis. The transit in‐vivo dosimetry performed by the electronic portal imaging device (EPID) avoids the problem of solid‐state detector positioning on the patient. Moreover, the dosimetric characterization of the recent Elekta aSi EPIDs in terms of signal stability and linearity make these detectors useful for the tran… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

1
9
0
3

Year Published

2011
2011
2019
2019

Publication Types

Select...
7

Relationship

2
5

Authors

Journals

citations
Cited by 12 publications
(13 citation statements)
references
References 15 publications
1
9
0
3
Order By: Relevance
“…The linac is equipped with Elekta IviewGT EPID, based on aSi panels XRD 1640 AL5 (PerkinElmer Optoelectronics, Fremont, CA, USA) operating at the fixed source-EPID distance (SED) equal to 160 cm. The IviewGT characteristics are discussed in [15]. An integrated transit signal St(x,y) can be obtained by the following equation:…”
Section: Equipment (Linac and Epid)mentioning
confidence: 99%
See 1 more Smart Citation
“…The linac is equipped with Elekta IviewGT EPID, based on aSi panels XRD 1640 AL5 (PerkinElmer Optoelectronics, Fremont, CA, USA) operating at the fixed source-EPID distance (SED) equal to 160 cm. The IviewGT characteristics are discussed in [15]. An integrated transit signal St(x,y) can be obtained by the following equation:…”
Section: Equipment (Linac and Epid)mentioning
confidence: 99%
“…In recent years, the authors developed a generalized procedure for the isocenter dose reconstruction of 3D-CRT treatments (Diso), based on the correlation functions between EPID transit signals and the doses at phantom mid-plane [11][12][13][14][15]. This research led to the foundation of an Italian national project financed by the Istituto Nazionale di Fisica Nucleare (INFN) and the Diso software was extensively applied in clinical routine in several Italian centers.…”
Section: Introductionmentioning
confidence: 99%
“…Further work has enabled this group to implement their in vivo dose-at-isocenter verification in various centers in Italy (29) and to apply it in numerous treatment situations. (30)(31)(32) We expanded the work by Piermattei et al to calculate 2D dose maps in an anthropomorphic phantom during three-dimensional conformal RT (3D CRT). To this aim, we had to characterize the off-axis variability of the correlation ratios F, here defined as the ratios of the EPID signals and the TPS doses in equivalent water phantom midplanes.…”
mentioning
confidence: 99%
“…At our institution, we developed a generalized procedure for the daily in‐vivo isocenter dose reconstruction of radiation treatments, leading to the foundation of an Italian national project financed by the Istituto Nazionale di Fisica Nucleare and to the development of the SOFTDISO software (Best Medical Italy, Italy). The analysed results for conformal radiotherapy reported clinically relevant differences between planned and delivered dose, detecting the presence of dose discrepancies in more than 10% of the tests with respect to our tolerance levels .…”
Section: Introductionmentioning