2023
DOI: 10.1002/anr3.12239
|View full text |Cite
|
Sign up to set email alerts
|

Contamination of the central medical air supply with water leading to mass ventilator failure

Abstract: SummaryHere, we present a case of mass ventilator failure due to contaminated medical air. Multiple ventilators failed routine tests, including almost all of the ventilators in our intensive care unit. A faulty air compressor had led to water contamination of our centre's supply of medical air. Water entered the pipeline supply of air and, hence the ventilators and anaesthetic machines. The disruption of the machines' proportional mixer valve resulted in unreliable delivery of fresh gas flow. This malfunction … Show more