1999
DOI: 10.1002/(sici)1522-2586(199901)9:1<119::aid-jmri16>3.0.co;2-f
|View full text |Cite
|
Sign up to set email alerts
|

Construction of a protocol for measuring blood flow by two-dimensional phase-contrast MRA

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
68
0
2

Year Published

1999
1999
2017
2017

Publication Types

Select...
8

Relationship

1
7

Authors

Journals

citations
Cited by 60 publications
(70 citation statements)
references
References 33 publications
0
68
0
2
Order By: Relevance
“…Flow measurements made in vivo are only approximations based on results of phantom studies (12). Therefore, the overall error of phasecontrast measurements is difficult to estimate.…”
Section: Errors and Limitationsmentioning
confidence: 99%
See 1 more Smart Citation
“…Flow measurements made in vivo are only approximations based on results of phantom studies (12). Therefore, the overall error of phasecontrast measurements is difficult to estimate.…”
Section: Errors and Limitationsmentioning
confidence: 99%
“…Accelerated flow (eg, turbulence or stenotic jets) causes the precision of the flow mea surement to decline (12), although this error might be corrected with a short echo time (34).…”
Section: Accelerated Flow and Spatial Misregistrationmentioning
confidence: 99%
“…One group of methods proposes estimating the velocity offset error in the vessel by examining the velocity offsets in somewhat distant stationary tissue (e.g., chest wall, liver, etc.) and spatially fitting the data using linear or higherorder interpolation (9)(10)(11). A drawback of this technique is that there may be insufficient stationary tissue to accurately determine the offset error in the vessel.…”
Section: Introductionmentioning
confidence: 99%
“…(1)(2)(3), the output of the PC MRI-like data generator is a set of images (or matrices) differing for slice thickness, in-plane pixel size and number of time frames per cardiac cycle.…”
Section: Pc Mri-like Data Generatormentioning
confidence: 99%
“…In vitro validation is limited by costs, time consumption, proper experimental setup to be built, intrinsic source of noise and/or artifacts such as eddy currents, concomitant gradients, displacement, misregistration, partial volume effects, intra-voxel phase dispersion and aliasing [1][2][3]26]. As for theoretical models [5,35], they remain extremely simplistic descriptors of cardiovascular flows.…”
mentioning