Considerations concerning the use of counting active personal dosemeters in pulsed fields of ionising radiation

Abstract: Active personal electronic dosimeters (APDs) exhibit limitations in pulsed radiation fields, which cannot be overcome without the use of new detection technology. As an interim solution, this paper proposes a method by which some conventional dosimeters can be operated in a way such that, based on the basic knowledge about the pulsed radiation field, any dosimetric failure of the dosimeter is signalised by the instrument itself. This method is not applicable to all combinations of APD and pulsed radiation fiel… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
6
0

Year Published

2010
2010
2024
2024

Publication Types

Select...
8
1

Relationship

0
9

Authors

Journals

citations
Cited by 23 publications
(6 citation statements)
references
References 2 publications
0
6
0
Order By: Relevance
“…Since all selected APDs, except PM1621A, provide a reading in the pulsed mode, this means they could be used in routine monitoring at hospitals, provided correction factors are introduced (11) . Some preliminary guidelines with recommendations on how to use APDs in practice were presented at IRPA2010 (12) .…”
Section: Resultsmentioning
confidence: 99%
“…Since all selected APDs, except PM1621A, provide a reading in the pulsed mode, this means they could be used in routine monitoring at hospitals, provided correction factors are introduced (11) . Some preliminary guidelines with recommendations on how to use APDs in practice were presented at IRPA2010 (12) .…”
Section: Resultsmentioning
confidence: 99%
“…In addition, it is recommended to use the audio or visual alarm, both in dose and dose rate regime. The dose and dose rate level to which an alarm would be set depends on the type of practice and characteristics of both the dosimeter and radiation field (Ambrosi et al 2010).…”
Section: Discussionmentioning
confidence: 99%
“…Given that accurately measuring the dose and dose rate in pulsed radiation fields is challenging for both the manufacturers and users of devices as well as organizations involved in licensing, this should be reviewed to maintain radiation safety [11][12]. Even though only technical specifications and articles regarding measurements in pulsed radiation fields are available, direct testing methods or standards are not approved yet by the Hungarian measurement office [13][14][15][16][17][18][19][20].…”
Section: Previous Measurements In Pulsed Radiation Fieldsmentioning
confidence: 99%