2014
DOI: 10.1136/neurintsurg-2014-011331
|View full text |Cite
|
Sign up to set email alerts
|

Compliant neurovascular balloon catheters may not be compatible with liquid embolic materials: intraprocedural rupture of the protecting balloon during tumor embolization using n-butyl cyanoacrylate and lipiodol mixture

Abstract: Glue embolization should not be performed with all three kinds of balloon catheter on the market, but DMSO and Onyx are compatible with those balloon catheters.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
2
0

Year Published

2018
2018
2023
2023

Publication Types

Select...
4
2

Relationship

0
6

Authors

Journals

citations
Cited by 7 publications
(2 citation statements)
references
References 6 publications
(4 reference statements)
0
2
0
Order By: Relevance
“…Among different LEAs, Onyx has become the mainstay treatment in recent years and has been shown to be equally, if not more, effective compared to n-BCA. 34 Despite reported incompatibility of n-BCA with DLBCs, 23,35 this practice has been endeavored in vivo by different interventionalists, without any reported intraoperative balloon rupture. 24,36…”
Section: Discussionmentioning
confidence: 99%
“…Among different LEAs, Onyx has become the mainstay treatment in recent years and has been shown to be equally, if not more, effective compared to n-BCA. 34 Despite reported incompatibility of n-BCA with DLBCs, 23,35 this practice has been endeavored in vivo by different interventionalists, without any reported intraoperative balloon rupture. 24,36…”
Section: Discussionmentioning
confidence: 99%
“…1,2 In addition, the ability to advance the catheter into distal small vessels and achieve flow control at the balloon level diminishes the risk of reflux proximally in the parent vessel, where essential branches might emerge from it. Even though the SMC is designed to be compatible only with ethylene vinyl alcohol (Onyx) and not with n-BCA, as the latter can cause rupture of the balloon and loss of flow control, 16,17 n-BCA has been used with the SMC in vivo, off label, in two VOGM cases 8,9 and in an oropharyngeal hemorrhage case which was controlled with lingual artery embolization. 9 Utilizing balloon microcatheters comes with risks inherent to their design.…”
Section: Discussionmentioning