Comparison of Vitros Dry Slide Technology for Determination of Lithium Ions with Other Methods

Abstract: The lithium ions concentration in human serum was determined using Dry-slide technology Vitros 250 Analyser (Ortho Clinical Diagnostic), atomic absorption spectrometry (AAS) method Perkin Elmer 403 and ion-selective electrode (ISE) potentiometry AVL 9181. We compared lithium ions results in sample sera between these methods. Our reference method was AAS. We analyzed lithium ions concentration in 23 sera samples of patients after oral administration of lithium carbonate (3x 300mg) Jadran, Galen Laboratory Rijek… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

0
5
0

Year Published

2016
2016
2023
2023

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(5 citation statements)
references
References 3 publications
(3 reference statements)
0
5
0
Order By: Relevance
“…They can detect Li + concentration as low as 10 μM, which is substantially lower than the concentration of Li + in non-invasively retrievable biofluids (e.g., saliva: 0.5-3 mM; ∼ double the corresponding serum concentrations). [23,24] Furthermore, Figure S2C (Supporting Information) shows that the sensor's response to the variations of interfering ions was minimal as compared to that of Li + , demonstrating good selectivity in the envisioned clinical application.…”
Section: Resultsmentioning
confidence: 99%
“…They can detect Li + concentration as low as 10 μM, which is substantially lower than the concentration of Li + in non-invasively retrievable biofluids (e.g., saliva: 0.5-3 mM; ∼ double the corresponding serum concentrations). [23,24] Furthermore, Figure S2C (Supporting Information) shows that the sensor's response to the variations of interfering ions was minimal as compared to that of Li + , demonstrating good selectivity in the envisioned clinical application.…”
Section: Resultsmentioning
confidence: 99%
“…From the literature, Serdarević N et al have compare the lithium values obtained from the ISE method with Vitros dry chemistry method and atomic spectrophotometry method, where they have mentioned that there was a positive interference in the ISE method. The Vitros dry slide technology method provides reliable lithium concentration results and is a valid alternative to Atomic Absorption Spectrometry method [14].…”
Section: Discussionmentioning
confidence: 99%
“…The VITROS dry slide technology method provides reliable Li concentration results and is a valid alternative to atomic absorption spectrometry method. 15 ►Table 2 explains the decreased value of Li in the clot activator vacutainers that may be due to the presence of silica additives in the silicone-coated vacutainers that on exposure of the Li by dry chemistry analyzer results in decreased Li values. Crown ethers are well known for binding metal cations due to their cyclic cavities and electron donor features.…”
Section: Discussionmentioning
confidence: 99%
“…The VITROS dry slide technology method provides reliable Li concentration results and is a valid alternative to atomic absorption spectrometry method. 15 …”
Section: Discussionmentioning
confidence: 99%