Comparison of Simulated and True Keratometry Measurements Using Swept-Source Optical Coherence Tomography and Dual Scheimpflug–Placido Imaging

Abstract: Purpose. To compare simulated and total keratometry and corneal astigmatism values between the IOLMaster 700 and Galilei G4 devices. Methods. A retrospective chart review was conducted for all patients undergoing phacoemulsification by a single surgeon (RTP) from March through September 2020 and who underwent imaging with both the IOLMaster 700 and Galilei G4. Exclusion criteria were prior corneal surgery, keratectatic diseases and inability to obtain a reliable image during image acquisition. Mean, flat, and … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2

Citation Types

0
6
0

Year Published

2023
2023
2024
2024

Publication Types

Select...
4

Relationship

1
3

Authors

Journals

citations
Cited by 4 publications
(9 citation statements)
references
References 15 publications
0
6
0
Order By: Relevance
“…These results are consistent with prior studies on normal eyes that showed highly correlated and interchangeable SimK values between the IOLM and G4 but with a consistently steeper True-K value by approximately 1.0 D on the IOLM vs the G4. 3,5 Lu et al demonstrated that posterior K values (mean, flat, and steep) were moderately correlated and not interchangeable between devices, with the IOLM measuring flatter posterior corneal curvatures, thus resulting in steeper True-K values with the IOLM. 5…”
Section: Discussionmentioning
confidence: 99%
See 4 more Smart Citations
“…These results are consistent with prior studies on normal eyes that showed highly correlated and interchangeable SimK values between the IOLM and G4 but with a consistently steeper True-K value by approximately 1.0 D on the IOLM vs the G4. 3,5 Lu et al demonstrated that posterior K values (mean, flat, and steep) were moderately correlated and not interchangeable between devices, with the IOLM measuring flatter posterior corneal curvatures, thus resulting in steeper True-K values with the IOLM. 5…”
Section: Discussionmentioning
confidence: 99%
“…These results are consistent with prior studies on normal eyes that showed highly correlated and interchangeable SimK values between the IOLM and G4 but with a consistently steeper True-K value by approximately 1.0 D on the IOLM vs the G4. 3,5 Lu et al demonstrated that posterior K values (mean, flat, and steep) were moderately correlated and not interchangeable between devices, with the IOLM measuring flatter posterior corneal curvatures, thus resulting in steeper True-K values with the IOLM. 5 Various reasons have been attributed to explain the difference in posterior corneal measurements and therefore true corneal power between SS-OCT and Scheimpflug corneal tomographers, such as the different hardware and technologies of the devices, as well as the software including different densities and locations of posterior corneal points measured between the devices.…”
Section: Discussionmentioning
confidence: 99%
See 3 more Smart Citations