J. Clin. Lab. Anal. volume 29, issue 5, P397-404 2014 DOI: 10.1002/jcla.21786 View full text
|
|
Share
Nihal Boğdaycioğlu, Fatma Meric Yilmaz, Sevilay Sezer, Esra Oğuz

Abstract: iSED sedimentation showed a poor correlation and a high bias (>10%) with the Westergren method, and Ves-Matic Cube 200 method showed a higher correlation and a lower bias than the iSED device when compared with Westergren reference method. These instruments should be carefully monitored.