2014
DOI: 10.1002/jcla.21786
|View full text |Cite
|
Sign up to set email alerts
|

Comparison of iSED and Ves‐Matic Cube 200 Erythrocyte Sedimentation Rate Measurements With Westergren Method

Abstract: iSED sedimentation showed a poor correlation and a high bias (>10%) with the Westergren method, and Ves-Matic Cube 200 method showed a higher correlation and a lower bias than the iSED device when compared with Westergren reference method. These instruments should be carefully monitored.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

11
29
1

Year Published

2016
2016
2023
2023

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 25 publications
(41 citation statements)
references
References 8 publications
11
29
1
Order By: Relevance
“…Hereby, we demonstrated a much better precision of iSED with commercial control samples at both evaluated levels than in a previously published study, while patient samples yielded intrarun CVs similar to already reported data for iSED, with decreasing precision at medium and low ESR levels This is in concordance with data from a lot of validation studies of automated ESR analyzers, regardless of the analyzer type and measurement principle and can be attributed to small numbers rather than poor analytical performance, therefore not being considered clinically significant.…”
Section: Discussionsupporting
confidence: 90%
See 4 more Smart Citations
“…Hereby, we demonstrated a much better precision of iSED with commercial control samples at both evaluated levels than in a previously published study, while patient samples yielded intrarun CVs similar to already reported data for iSED, with decreasing precision at medium and low ESR levels This is in concordance with data from a lot of validation studies of automated ESR analyzers, regardless of the analyzer type and measurement principle and can be attributed to small numbers rather than poor analytical performance, therefore not being considered clinically significant.…”
Section: Discussionsupporting
confidence: 90%
“…The comparison between iSED and the Westergren method revealed higher correlation coefficients than in a previous validation study. 13 As evidenced in this study, iSED analyzer provides accurate ESR measurement and acceptable concordance with the gold standard Westergren method, with minor discrepancies and occasionally limited agreement that can be attributed to different methodological principles. However, as ESR could be considered more a generic sign of pathophysiological alterations than a true laboratory measure, not particularly dense of specific clinical significance, for this target, iSED could provide acceptable discriminatory performance.…”
Section: Discussionmentioning
confidence: 51%
See 3 more Smart Citations