Comparison of an anthropomorphic PRESAGE® dosimeter and radiochromic film with a commercial radiation treatment planning system for breast IMRT: a feasibility study

Abstract: This work presents a comparison of an anthropomorphic PRESAGE® dosimeter and radiochromic film measurements with a commercial treatment planning system to determine the feasibility of PRESAGE® for 3D dosimetry in breast IMRT. An anthropomorphic PRESAGE® phantom was created in the shape of a breast phantom. A five‐field IMRT plan was generated with a commercially available treatment planning system and delivered to the PRESAGE® phantom. The anthropomorphic PRESAGE® was scanned with the Duke midsized optical CT … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

2
22
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
6

Relationship

3
3

Authors

Journals

citations
Cited by 14 publications
(24 citation statements)
references
References 25 publications
2
22
0
Order By: Relevance
“…The maximum per cent dose difference was 4% between Oncentra ® and PRESAGE ® . Preliminary investigations suggest the cause may be a reflection artefact of laser light from the underside of the top and bottom of the dosimeter 32 Figure 5(a) Comparison of the calculated and measured point skin doses for PRESAGE ® , Oncentra ® Brachy Planning TPS and EBT2 film for film 1.…”
Section: Resultsmentioning
confidence: 99%
“…The maximum per cent dose difference was 4% between Oncentra ® and PRESAGE ® . Preliminary investigations suggest the cause may be a reflection artefact of laser light from the underside of the top and bottom of the dosimeter 32 Figure 5(a) Comparison of the calculated and measured point skin doses for PRESAGE ® , Oncentra ® Brachy Planning TPS and EBT2 film for film 1.…”
Section: Resultsmentioning
confidence: 99%
“…The PRESAGE ® PTV DVH indicates that the delivered dose was somewhat less homogeneous than that calculated using the Pinnacle 3 treatment planning system, with a maximum of 2.2% of small regions of the relationship above and below the dose occurring near the edge of the breast phantom. The average difference in the target volume was 1.5% with a standard deviation of 0.49%, and a part of this difference is real, whereas some of it was due to artefacts in the PRESAGE ® distribution 15 …”
Section: Discussionmentioning
confidence: 93%
“…Scans normalised the calculated dose distribution and 3D comparison was made with a treatment planning system 16 . The DMOS-RPC scanner shown in Figure 1 (left) consists of a matched telemetry source and image lenses provide visual field 15 . In graphical user interface, the top-left quadrants of projection screen images are associated with painting on preliminary irradiation, and the bottom-left projection screens are associated with the image scan after irradiation.…”
Section: Methodsmentioning
confidence: 99%
See 2 more Smart Citations