Comparison of air kerma measurements between the PTB and the IAEA for x-radiation qualities used in general diagnostic radiology and mammography

Abstract: A comparison of the air kerma standards for x-radiation qualities used in general diagnostic radiology and mammography, identified as EURAMET.RI(I)-S10 (EURAMET project #1221), was performed between the PTB and the IAEA. Two spherical and two parallel-plate reference-class ionization chambers of the IAEA and 12 beam qualities standardized in the IEC standard 61267:2005 plus additional 7 standard beam qualities established at both laboratories were selected for the comparison. The calibration coefficients were … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
8
0

Year Published

2015
2015
2018
2018

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 7 publications
(8 citation statements)
references
References 4 publications
0
8
0
Order By: Relevance
“…Uncertainty of the air kerma rate determination with the PK100 was 0.3%, uncertainty on monitor normalization was 0.1%, and uncertainty on beam nonuniformity was 0.1%. 36 Uncertainty of the calibration in terms of D w was 1.1%, 0.9%, 0.8%, 0.7%, 0.7%, and 0.6% for T70, T100, T150, T200, T250, and T 280, respectively. 37 Reproducibility of measurements at the 90 Sr source has been estimated to be 0.6%, and this uncertainty has been assigned to current measurements at the 90 Sr source.…”
Section: D Uncertaintiesmentioning
confidence: 91%
“…Uncertainty of the air kerma rate determination with the PK100 was 0.3%, uncertainty on monitor normalization was 0.1%, and uncertainty on beam nonuniformity was 0.1%. 36 Uncertainty of the calibration in terms of D w was 1.1%, 0.9%, 0.8%, 0.7%, 0.7%, and 0.6% for T70, T100, T150, T200, T250, and T 280, respectively. 37 Reproducibility of measurements at the 90 Sr source has been estimated to be 0.6%, and this uncertainty has been assigned to current measurements at the 90 Sr source.…”
Section: D Uncertaintiesmentioning
confidence: 91%
“…It is expressed as the difference of the companion result and the respective CRV. When a number of radiation qualities and instruments are used in a comparison, the results are deduced for each quality and instrument, separately [2,3,4,17,20], as described in the previous sections.…”
Section: Methods For Deriving the Degree Of Equivalencementioning
confidence: 99%
“…Twenty two (22) laboratories (PSDLs and SSDLs) participated in this EURAMET #1177 comparison (Table 1) The ICRL/GAEC-EIM, Greece (EL) as the pilot laboratory was responsible for the overall coordination of the comparison and the analysis and the reporting of the results.…”