2017
DOI: 10.1016/j.ejmp.2017.09.124
|View full text |Cite
|
Sign up to set email alerts
|

Comparison between new-generation SiPM-based and conventional PMT-based TOF-PET/CT

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
89
1
1

Year Published

2019
2019
2023
2023

Publication Types

Select...
6
1
1

Relationship

1
7

Authors

Journals

citations
Cited by 84 publications
(91 citation statements)
references
References 41 publications
0
89
1
1
Order By: Relevance
“…The NEMA NU 2-2012 parameters measured for the 4-ring DMI system (AFOV = 20 cm) were compared with results obtained at Stanford and Uppsala Universities [2] and Tokyo [3] for the same system and with results from Brugge [4] obtained for the system in its 3-ring configuration (AFOV = 15 cm). For the DMI-DR system, results were compared with those obtained at Southampton-Poole-Plymouth [5].…”
Section: Comparisons With Other Studiesmentioning
confidence: 99%
See 2 more Smart Citations
“…The NEMA NU 2-2012 parameters measured for the 4-ring DMI system (AFOV = 20 cm) were compared with results obtained at Stanford and Uppsala Universities [2] and Tokyo [3] for the same system and with results from Brugge [4] obtained for the system in its 3-ring configuration (AFOV = 15 cm). For the DMI-DR system, results were compared with those obtained at Southampton-Poole-Plymouth [5].…”
Section: Comparisons With Other Studiesmentioning
confidence: 99%
“…It is noteworthy that the results for the two systems are similar with, for instance, a spatial resolution for DMI and DMI-DR of 4.62 vs 4.58 for FBP, 3.85 vs 4.13 for VPHD (3D OSEM) and 2.70 vs 2.84 for VPHD-S (3D OSEM + PSF) at 1-cm radial offset. For the DMI evaluation, the results obtained at our institution are compared with previous results from Stanford and Uppsala Universities [2], Tokyo [3] and Brugge [4] in Fig. 1a-c.…”
Section: Spatial Resolutionmentioning
confidence: 99%
See 1 more Smart Citation
“…As a result, these detectors have seen over a decade of engineering 25,26 and broad applications in e.g. lidar 27 and medical imaging 28 .…”
Section: Introductionmentioning
confidence: 99%
“…The physical performance of two PET systems [9], one based on new-generation detectors of digital silicon photomultipliers (SiPM) and the other one with PMT and TOF, were evaluated using NEMA NU 2 (published by the National Electrical Manufacturers Association (NEMA)), provided by the vendor for each PET/CT scanner, allowing reproducible and accepted measurement standards for evaluating the physical performance of PET systems [10]. The SiPM-PET/CT demonstrated higher sensitivity, better performance under high count-rate conditions and better image quality than the conventional PMT-PET/CT system.…”
mentioning
confidence: 99%