Comparative Accuracy Analysis of a Real-time and an Intermittent-Scanning Continuous Glucose Monitoring System

Abstract: Background: Currently, two different types of continuous glucose monitoring (CGM) systems are available: real time (rt) CGM systems that continuously provide glucose values and intermittent-scanning (is) CGM systems. This study compared accuracy of an rtCGM and an isCGM system when worn in parallel. Methods: Dexcom G5 Mobile (DG5) and FreeStyle Libre (FL) were worn in parallel by 27 subjects for 14 days including two clinic sessions with induced glucose excursions. The percentage of CGM values within ±20% or ±… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
4
0

Year Published

2020
2020
2024
2024

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 14 publications
(4 citation statements)
references
References 14 publications
0
4
0
Order By: Relevance
“…A study of 72 diabetic patients evaluated a Dexcom G4 Platinum CGMS with a MARD value of 13% ( 10 ). The study on Dexcom G5 Platinum CGMS indicated a MARD value of 9.5% ( 11 ). In addition, a separate study of Dexcom G6 Platinum CGMS showed a MARD value of 9.0% ( 12 ).…”
Section: Discussionmentioning
confidence: 99%
“…A study of 72 diabetic patients evaluated a Dexcom G4 Platinum CGMS with a MARD value of 13% ( 10 ). The study on Dexcom G5 Platinum CGMS indicated a MARD value of 9.5% ( 11 ). In addition, a separate study of Dexcom G6 Platinum CGMS showed a MARD value of 9.0% ( 12 ).…”
Section: Discussionmentioning
confidence: 99%
“…It was previously shown that the FreeStyle Libre deviates systematically by -0.5 mmol/L in people with type 1 diabetes using HemoCue capillary measurements as a reference[ 15 ]. The Dexcom G5 was found to have a MARD of 7.1%-15.7% when tested in people with type 1 diabetes and using a Yellow Spring Instrument as a reference[ 16 ].…”
Section: Discussionmentioning
confidence: 99%
“…This feature allowed caretakers to remotely monitor glucose levels and offer treatment support. The same year, the G5 sensor was released, which incorporated wireless rapid data downloads on a smart phone and/or receiver 61 . The G6 was approved in 2018 as the first iCGM that allowed integration with different insulin pumps, such as Tandem and Omnipod 5 62 , 63 .…”