2017
DOI: 10.1016/j.ejmp.2017.01.005
|View full text |Cite
|
Sign up to set email alerts
|

Commissioning Monte Carlo algorithm for robotic radiosurgery using cylindrical 3D-array with variable density inserts

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

0
8
0

Year Published

2017
2017
2019
2019

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 13 publications
(8 citation statements)
references
References 29 publications
0
8
0
Order By: Relevance
“…Differences were between 3.6% and 5.1% for cones with diameter <10 mm. The ArcCHECK was recently investigated for commissioning of a Multiplan ® Monte Carlo dose calculation algorithm 44. It was found that while the ArcCHECK addresses some of the small‐field dosimetry challenges (its diodes have real‐time response, high sensitivity and sub‐mm lateral size of the SVs), the measurement of field sizes with diameter inferior or equal to the SVs pitch should be considered with care.…”
Section: Discussionmentioning
confidence: 99%
“…Differences were between 3.6% and 5.1% for cones with diameter <10 mm. The ArcCHECK was recently investigated for commissioning of a Multiplan ® Monte Carlo dose calculation algorithm 44. It was found that while the ArcCHECK addresses some of the small‐field dosimetry challenges (its diodes have real‐time response, high sensitivity and sub‐mm lateral size of the SVs), the measurement of field sizes with diameter inferior or equal to the SVs pitch should be considered with care.…”
Section: Discussionmentioning
confidence: 99%
“…This material assignment method could lead to incorrect simulation of physical interactions for high‐density metal. Since different materials of similar density could introduce differences in deposited dose, we should be cautious when working at the end of a material's density range 6, 20…”
Section: Discussionmentioning
confidence: 99%
“…But this commissioning was performed for their own dual‐source model built previously 5. Dechambre et al once proposed a novel method for commissioning the MC dose calculation algorithm of Multiplan using a cylindrical 3D‐array with variable density inserts 6. However, the diode spacing of these arrays (0.5–1.0 cm) is generally too large for the small field of Cyberknife, thus may not provide sufficient information for dose verification.…”
Section: Introductionmentioning
confidence: 99%
“…In addition to its application in dosimetric verification, its unique 3D design also makes it a potential tool for machine QA . Although the ArcCHECK has been widely used for QA of isocentric treatment machines, its applications in commissioning of Monte Carlo algorithm and patient‐specific QA for nonisocentric treatment machine such as CyberKnife were also studied by some groups . In this study, we developed a novel method for testing the constancy of beam path accuracy of our CyberKnife (model M6) system by using ArcCHECK.…”
Section: Introductionmentioning
confidence: 99%